ECscan20/30/40 

 

Appendix 

 

Preparation of Conductivity Standard Solutions 

1. 

Place the analytical grade potassium chloride (KCl) reagent in a beaker and dry in an oven for about 3 hours at 105° C/221° F, then cool to 
room temperature. 

2.    Add the reagent to a 1 litre volumetric flask according to the instructions in table below. 
3. 

Fill the deionized water to the mark, mix the solution until the reagent is completely dissolved. 

 

Conductivity Standard 

Reagent 

Weight 

84

 

µ S/cm 

KCl 

42.35

 

mg 

1413

 

µ S/cm 

KCl 

745.9

 

mg 

12.88

 

mS/cm 

KCl 

7.45

 

mg 

 

Calculating the TDS Conversion Factor 

The following formula describes the calculation method of TDS conversion factor. 
 
 
 

Factor =   

 

 
 
 

Where: 

Actual TDS = Value from the high purity water and precisely weighed NaCl or KCL reagent. 

 

Actual Conductivity = The meter measured conductivity value. 

 
For example, dissolve 64 grams of KCl reagent in 1 litre deionized water. If measured conductivity value is 100

 

mS/cm, then TDS factor is 0.64. 

 

Conductivity to TDS Conversion Factors 

 

Conductivity at 25° C 

TDS (KCl) 

TDS (NaCl) 

ppm 

Factor 

ppm 

Factor 

84

 

µ S/cm 

40.38 

0.5048 

38.04 

0.4755 

1413

 

µ S/cm 

744.7 

0.527 

702.1 

0.4969 

12.88

 

mS/cm 

7447 

0.5782 

7230 

0.5613 

 
Optional Accessories 

 
Order Code 

Description 

E-ECscan-C1-10K 

Platinum conductivity electrode, cell constant 1, measurement range: 10

 

μ

S/cm to 20

 

mS/cm

 

E-ECscan-C10-10K 

Platinum conductivity electrode, cell constant 10, measurement range: 100

 

μ

S/cm to 200

 

mS/cm 

ECCS-84 

Conductivity standard solution 146.5

 

μ

S/cm, 480

 

ml 

ECCS-1413 

Conductivity standard solution 1413

 

μ

S/cm, 480

 

ml 

ECCS-1288 

Conductivity standard solution 12.88

 

mS/cm, 480

 

ml 

 
 
 
 
 

Actual TDS

 

Actual Conductivity @25° C

 

Содержание ECscan40

Страница 1: ...ECscan20 30 40 Pocket Conductivity Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...urement Exit the calibration setting and return to the measurement mode Start calibration Press and hold key to enter the setup menu Select an option Confirm the calibration setting or displayed optio...

Страница 3: ...Disable When the option is enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Enable Disable Reset the tester to factory default settings Enable Disable Disa...

Страница 4: ...ster enters the calibration mode the display shows CAL 1 1 2 Rinse the electrode with deionized water and place into the standard solution Stir tester gently to remove air bubbles trapped in the slot...

Страница 5: ...he Enter the tester will show Conductivity TDS Salinity icon and switch to the corresponding measurement mode Measuring the Sample Rinse the electrode with deionized water Place the electrode into the...

Страница 6: ...precisely weighed NaCl or KCL reagent Actual Conductivity The meter measured conductivity value For example dissolve 64 grams of KCl reagent in 1 litre deionized water If measured conductivity value i...

Страница 7: ...ndensing Display Dual line LCD 21 21mm 0 82 0 82 Power Requirements 2 1 5V AAA alkaline batteries Auto Off 8 minutes after last key pressed Dimensions 185 L 40 Dia mm 7 28 1 57 Weight 100g 3 5oz Dispo...

Страница 8: ...ce 4715 Castlewood St Sugar land TX 77479 USA Tel 1 346 762 7358 E mail banteinstruments yahoo com Factory A10 No 2066 Laifang Rd Shanghai 201615 China Tel 86 21 6404 1598 E mail banteinstrument hotma...

Отзывы: