Bante Instruments ECscan40 Скачать руководство пользователя страница 5

ECscan20/30/40 

 

Temperature Calibration 

The

 

ECscan series tester has a built-in temperature sensor that used for automatic temperature compensation (ATC). Durning the measurement, 

if the measured temperature reading differs from that of an accurate thermometer, the tester needs to be calibrated. 
1.    Press and hold the 

Cal

 key to enter the setup menu. 

2. 

Press the 

Enter

 key until the display shows

 

/

 

 

or

 

/

 

3. 

Press the 

Cal

 key, the tester enters the temperature calibration mode. 

4. 

Place the electrode into a solution with a known accurate temperature and wait for measurement is stable.   

5. 

Press the 

Cal

 key to modify the temperature value. 

6. 

Press the 

Enter

 key to save and press the 

Meas

 key to return to the measurement mode. 

 

 

 

 

 

Measurements 

 

Switching the Measurement Mode 

Press the 

Enter

, the tester will show 

 

(Conductivity), 

 

(TDS), 

 

(Salinity) icon and switch to the corresponding measurement 

mode. 
 

Measuring the Sample 

Rinse the electrode with deionized water. Place the electrode into the sample solution and stir gently. Make sure that no air bubbles on the sensor 
surface. Wait for the measurement to stabilize and record the reading. 
 

   

 

If the display shows "----" indicating the measurement exceeds the range, remove the tester from the sample immediately. 

 

If the

 

 

option is enabled in the setup menu, the tester will automatically lock a measurement endpoint and show

 

 

icon. 

 

Press the 

Meas

 key to resume measuring. 

 

Electrode Maintenance 

 

Rinse the electrode thoroughly with deionized water after use. 

 

Do not touch the platinum black coating on sensor surface and always keep it clean. 

 

If there is a build-up of solids inside the sensor, remove very carefully, then recalibrate the tester. 

 

If you do not use the tester for long periods, store the electrode with tap water. 

 
 
 

 

Содержание ECscan40

Страница 1: ...ECscan20 30 40 Pocket Conductivity Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...urement Exit the calibration setting and return to the measurement mode Start calibration Press and hold key to enter the setup menu Select an option Confirm the calibration setting or displayed optio...

Страница 3: ...Disable When the option is enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Enable Disable Reset the tester to factory default settings Enable Disable Disa...

Страница 4: ...ster enters the calibration mode the display shows CAL 1 1 2 Rinse the electrode with deionized water and place into the standard solution Stir tester gently to remove air bubbles trapped in the slot...

Страница 5: ...he Enter the tester will show Conductivity TDS Salinity icon and switch to the corresponding measurement mode Measuring the Sample Rinse the electrode with deionized water Place the electrode into the...

Страница 6: ...precisely weighed NaCl or KCL reagent Actual Conductivity The meter measured conductivity value For example dissolve 64 grams of KCl reagent in 1 litre deionized water If measured conductivity value i...

Страница 7: ...ndensing Display Dual line LCD 21 21mm 0 82 0 82 Power Requirements 2 1 5V AAA alkaline batteries Auto Off 8 minutes after last key pressed Dimensions 185 L 40 Dia mm 7 28 1 57 Weight 100g 3 5oz Dispo...

Страница 8: ...ce 4715 Castlewood St Sugar land TX 77479 USA Tel 1 346 762 7358 E mail banteinstruments yahoo com Factory A10 No 2066 Laifang Rd Shanghai 201615 China Tel 86 21 6404 1598 E mail banteinstrument hotma...

Отзывы: