ECscan20/30/40 

 

Switching the Tester On and Off 

 

Press and hold the 

Meas

 key for about 5 seconds to switch on the tester. 

 

Press and hold the 

Meas

 key to switch off the tester. 

 

To enable the auto-power off function, refer to the 

Setup Menu

 section.

 

 

Setup Menu 

The ECscan series tester contains 7 menu items in the setup menu, the following table describes the functions of each option. 
 

Menu 

Description 

Options 

Description 

Default 

 

Set the number of calibration points 

 

1, 2 or 3 points 

1 point 

 
 

 

Set the TDS conversion factor 

 

Range: 0.1 to 1.0 (only for 
ECscan30 and 40 testers) 

0.5 

 

Set the temperature unit 

 

Degrees Celsius 

° C 

 

Degrees Fahrenheit 

 

Calibrate the temperature 

 

Reading ± 10° C/° F 

--- 

 

 

When the option is enabled, the tester will automatically   
sense a stable reading and lock the measurement 

 

Enable 

Disable 

 

Disable 

 

When the option is enabled, the tester will automatically   
switch off if no key is pressed within 8 minutes 

 

Enable 

Enable 

 

Disable 

 

Reset the tester to factory default settings 

 

Enable 

Disable 

 

Disable 

 

Setting the Default Option 

1. 

In the measurement mode, press and hold the 

Cal

 key for 5 seconds to enter the setup menu. 

2. 

If necessary, press the 

Cal

 key again to select an option. 

3. 

Press the 

Enter

 key, the tester saves the current option and moves to the next menu item. 

4. 

Repeat the steps above until the tester returns to the measurement mode. Setting is completed. 

 

 

 

 

 

 

During the setting process, press the 

Meas

 key, the tester will exit the setup menu and return to the measurement mode. 

 

If you do not need to calibrate the temperature, press the 

Enter

 key to skip the

 

/

 

 

or

 

/

 

 

option. 

 

The

 

 

option is used to restore the tester back to the factory default settings. If enabled, all of the calibration data and user-specific 

 

settings will be deleted or reset, the tester must be recalibrated. 

 

option 

menu 
item 

Содержание ECscan20

Страница 1: ...ECscan20 30 40 Pocket Conductivity Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...urement Exit the calibration setting and return to the measurement mode Start calibration Press and hold key to enter the setup menu Select an option Confirm the calibration setting or displayed optio...

Страница 3: ...Disable When the option is enabled the tester will automatically switch off if no key is pressed within 8 minutes Enable Enable Disable Reset the tester to factory default settings Enable Disable Disa...

Страница 4: ...ster enters the calibration mode the display shows CAL 1 1 2 Rinse the electrode with deionized water and place into the standard solution Stir tester gently to remove air bubbles trapped in the slot...

Страница 5: ...he Enter the tester will show Conductivity TDS Salinity icon and switch to the corresponding measurement mode Measuring the Sample Rinse the electrode with deionized water Place the electrode into the...

Страница 6: ...precisely weighed NaCl or KCL reagent Actual Conductivity The meter measured conductivity value For example dissolve 64 grams of KCl reagent in 1 litre deionized water If measured conductivity value i...

Страница 7: ...ndensing Display Dual line LCD 21 21mm 0 82 0 82 Power Requirements 2 1 5V AAA alkaline batteries Auto Off 8 minutes after last key pressed Dimensions 185 L 40 Dia mm 7 28 1 57 Weight 100g 3 5oz Dispo...

Страница 8: ...ce 4715 Castlewood St Sugar land TX 77479 USA Tel 1 346 762 7358 E mail banteinstruments yahoo com Factory A10 No 2066 Laifang Rd Shanghai 201615 China Tel 86 21 6404 1598 E mail banteinstrument hotma...

Отзывы: