Appendix
I
8
Meter Specifications
Model
Bante 510
Conductivity
Range
0.01 µ S/cm to 200.0 mS/cm
Resolution
0.001, 0.01, 0.1, 1
Accuracy
± 1% F.S.
Calibration Points
1 to 3 points
Calibration Solutions
10 µ S/cm, 84 µ S/cm, 1413 µ S/cm,
12.88 mS/cm, 111.8 mS/cm
Temperature Compensation
0 to 100° C (32 to 212° F), manual or
automatic
Temperature Coefficient
0.0 to 9.9%/° C
Reference Temperature
25° C
Cell Constant
K = 0.1, 1, 10 or custom
Temperature
Range
0 to 105° C (32 to 221° F)
Resolution
0.1° C (0.1° F)
Accuracy
± 1° C (± 1.8° F)
Calibration Point
1 point
Other Specifications
Operating Temperature
0 to 50° C (32 to 122° F)
Storage Temperature
0 to 60° C (32 to 140° F)
Relative Humidity
< 80% (non-condensing)
Display
LCD, 135 × 75 mm (5.3 × 2.9 in.)
Power Requirements
9V/400mA DC power adapter
Auto-Off
3 hours after last key pressed
Dimensions
210 (L) × 205 (W) × 75 (H) mm,
(8.2 × 8.0 × 2.9 in.)
Weight
1.5 kg (3.3 lb)
Troubleshooting
Fault
Cause and Corrective Action
Electrode dried out.
Soak the conductivity electrode in tap water for
about 10 minutes.
Measurement exceeded the maximum range.
Check the electrode and sample.
Drifting erratic
readings
Check whether electrode is contaminated, clogged
or broken.
Setting value does not match calibration solution.
Reset the calibration value or check the calibration
solution.
Electrode is broken.
Replace the conductivity electrode.