Bante932 Benchtop Water Hardness Meter
11
1.5
Wait for the reading to stabilize, the display shows 0.1mmol/L/CAL2. The meter prompts you
to continue with second point calibration.
1.6
Rinse the water hardness electrode with distilled water, then rinse with a small amount of
standard solution.
Place the electrode into the next standard solution.
Stir the electrode gently.
Press the
Enter
key, the Calibration icon begins flashing.
1.7
Wait for the reading to stabilize, the meter automatically show END and return to the
measurement mode. Calibration is completed.
•
If you have selected the multi-point calibration in the setup menu, the display will show CAL3.
The meter prompts you to continue with third point calibration. Repeat the step 1.6 above
until the display shows END. The meter will automatically return to the measurement mode.
•
If you want to exit the calibration, press the
Meas
key.