Bante Instruments Bante932 Скачать руководство пользователя страница 12

Bante932 Benchtop Water Hardness Meter 

11 

 

 

   
1.5 

Wait for the reading to stabilize, the display shows 0.1mmol/L/CAL2. The meter prompts you 

 

to continue with second point calibration. 

 

   
1.6 

Rinse the water hardness electrode with distilled water, then rinse with a small amount of 

 

standard solution.

 

Place the electrode into the next standard solution.

 

Stir the electrode gently. 

 

Press the 

Enter

 key, the Calibration icon begins flashing. 

 

   
1.7 

Wait for the reading to stabilize, the meter automatically show END and return to the   

  measurement mode. Calibration is completed. 

 

   

 

If you have selected the multi-point calibration in the setup menu, the display will show CAL3. 

 

The meter prompts you to continue with third point calibration. Repeat the step 1.6 above   

 

until the display shows END. The meter will automatically return to the measurement mode. 

 

If you want to exit the calibration, press the 

Meas

 key. 

 

 

 

 

 

 

 

 

 

 

 

Содержание Bante932

Страница 1: ...Bante932 Benchtop Water Hardness Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...t work environment meets following conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list...

Страница 3: ...meter is in the measurement mode Stable icon Indicates the measuring value has stabilized Calibration mode icon Indicates the meter is in the calibration mode Hold icon Indicates the measuring value...

Страница 4: ...menu Press and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down thr...

Страница 5: ...tion Adjustment of electrode arm After installation if the electrode arm automatically rises or falls you need to adjust the screws until arm locate at any position 1 Remove the plastic cover from the...

Страница 6: ...O NOT pull on the cable Always make sure that the connector is clean and dry Connecting the Temperature Probe 1 Place the temperature probe into the circular hole of the electrode arm 2 Insert the pho...

Страница 7: ...Use Remove the protective cap from the bottom of the water hardness electrode Soak the electrode in the 0 01mol L standard solution for at least 30 minutes Switching the Meter On and Off Press the Me...

Страница 8: ...Hold When the option is enabled the meter will automatically sense a stable reading and lock the measurements Enable Disable Disable Auto Power Off When the option is enabled the meter will automatica...

Страница 9: ...ter returns to the measurement mode Setting is completed If you want to exit the setting press the Meas key Setting the date and time 2 1 Press and hold the key for 3 seconds to enter the setup menu a...

Страница 10: ...meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indicating tha...

Страница 11: ...ted or calibration is not successfully the display will always show 0 000 During the calibration ensure that the selected calibration points cover the anticipated range of the samples Calibrating the...

Страница 12: ...Stir the electrode gently Press the Enter key the Calibration icon begins flashing 1 7 Wait for the reading to stabilize the meter automatically show END and return to the measurement mode Calibratio...

Страница 13: ...key until the meter shows ELE P 02 2 3 Press the Enter key the meter shows the last calibration date Format month day 2 4 Press the key the meter shows the first calibartion point and mV value e g 0...

Страница 14: ...ardness Measurement Before measuring ensure that the temperature of samples are the same as the standard solutions the maximum error should be controlled within the 1 C For low level measurements or s...

Страница 15: ...e display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 01 Data Log 2 Press the Enter key the meter shows the serial number of the stored data 3 Press the ke...

Страница 16: ...shed Click the OK button to confirm Click the Receive button the stored data automatically transfer to computer Interval recording This function is used for recording the measuring value within the sp...

Страница 17: ...ed value is out of range Check the electrode whether clogged dirty or broken Incorrect calibration solutions Using the fresh calibration solutions for calibration Electrode has expired Replace the ele...

Страница 18: ...on 0 50 C Manual or Automatic Memory Stores up to 500 data sets Output USB communication interface Connector BNC Display LCD Operating Temperature 0 60 C Relative Humidity 80 Power Requirements DC5V u...

Страница 19: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: