BANTE221 PORTABLE PH/ORP METER
5
Setup Menu
Bante221 portable meter contains an integrated setup menu that allows you to customize each displayed option to meet measurement
requirements.
pH MODE:
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
pH Buffers:
Select your desired pH buffers for calibration and
auto-recognition
USA Standard
■
NIST Standard
DIN Standard
User-Defined Buffer
Calibration Points:
Select the number of calibration points
1 point
2 points
3 points
■
4 points
5 points
Resolution:
Sets the resolution for pH measurement
0.001pH
■
0.01pH
0.1pH
Measurement Unit:
Sets the default temperature unit
Degrees Celsius
■
Degrees Fahrenheit
ORP MODE:
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Resolution:
Sets the resolution for ORP measurement
0.1mV
■
1mV
GENERAL OPTIONS:
PARAMETER
DESCRIPTION
OPTIONS
DESCRIPTION
DEFAULT
Stable Criteria:
Sets the stability standard for measurement or
calibration.
When the “LO” option is enabled, measuring
value will stabilize quickly.
When the “HI” option is
enabled, measuring value will stabilize slowly.
Low
■
High