Manual Grinding Wheel 2928 00

 Problems and solutions

 

 In the following chapter possible causes for faults and work needed for their

 

  resolution are described.

 

 For problems and faults that cannot be resolved using the following instructions,

 

  contact your authorised dealer or the manufacturer. For Germany see the service 

address on page 2.

9.1   Troubleshooting guide

 

 Personnel: Specialized personnel

Fault description

Cause

Remedy

Motor does not 

turn

Instrument

switched off

Switch on the device (Chapter 7.2.4 ‘Swit-

ching the instrument on/off’ on page 22).

Instrument not 

connected to the 

power supply

Connect to the power supply (Chapter 6.5 

‘Connect the unit to the power supply’ on 

page 19).

Power failure

Check the fuse of the power supply of the 

healthcare establishment.

Defective instru-

ment fuse

Replace the instrument fuse on the back of 

the instrument in the instrument inlet recep-

tacle fuse holder.

Instrument inlet

connector loose

Check instrument inlet connector is plugged 

firmly in.

Grinding speed is

inadequate

Grinding wheel

blocked up with

grinding material

Remove material from grinding wheel with the 

cleaning stone (Chapter 8.4 ‘Cleaning the 

grinding wheel (removal)’ on page 26).

Grinding wheel is 

dry although tap 

water is dripping 

from the water 

valve

Sponge is 

blocked up with 

grinding sludge

Clean the sponge or replace if necessary 

(Chapter 8.3 ‘Cleaning sponges’ on page 

24).

Water is running 

down the sides of 

the upper outer 

housing if tap 

water is dripping 

from the water 

valve

Sponge is 

blocked up with 

grinding sludge

Clean the sponge or replace if necessary 

(Chapter 8.3 ‘Cleaning sponges’ on page 

24).

Water leaks from 

the bottom of the 

manual grinding 

wheel

Collection contai-

ner is full

Empty the collection container (Chapter 7.2.1 

‘Check water collection container’ on page 

20).

Overflow hose 

on the collec-

tion container is 

kinked or blocked

Check the position of the overflow hose on 

the operator‘s side or remove the blockage.

Problems and solutions

Содержание 2928 00

Страница 1: ...Operating instructions Read the instructions prior to performing any task Hand grinding stone with and without sensor 2928 00 2928 10...

Страница 2: ...Planck Stra e 30 61184 Karben Germany Telephone 49 0 60 39 993 10 Fax 49 0 60 39 993 37 email info b s de Internet www b s de 03 2014 These operating instructions were written by Breitfeld Schliekert...

Страница 3: ...on 13 4 1 Overview of the manual grinding wheel 13 4 2 Power switch 13 4 3 Direction of rotation switch 14 4 4 Instrument inlet receptacle with fuse storage 14 4 5 Power supply cord 14 4 6 Water colle...

Страница 4: ...4 8 3 1 Cleaning the top sponge 24 8 3 2 Cleaning the bottom sponge 25 8 4 Cleaning the grinding wheel removal 26 8 5 Changing the grinding wheel 27 9 Problems and solutions 28 9 1 Troubleshooting gui...

Страница 5: ...tructions and use caution to prevent accidents personal injury and material damage DANGER DANGER This combination of symbol and keyword points to a situa tion of immediate danger which may lead to ser...

Страница 6: ...result of the latest technical modifications The undertakings agreed in the supply contract as well as the manufacturer s Terms and Conditions and Terms of Delivery and the legal regulations applicab...

Страница 7: ...eep untrained persons away from the instrument Do not set up the instrument near optical measuring instruments Any claims due to damage arising from incorrect use are excluded 2 2 Responsibility of th...

Страница 8: ...he applicable standards and regula tions are capable of executing the tasks assigned to them and of recognizing possible hazards and avoiding them on their own Only skilled personnel are approved as p...

Страница 9: ...h extreme caution All work must only be done by skilled personnel NOTICE NOTICE The instrument may be damaged by incorrect use Improper use may damage the instrument Do not drop the instrument Do not...

Страница 10: ...must not be disposed of in household rubbish The obsolete instrument can be sent to waste management for electronics and metal recycling The manufacturer also accepts old instruments for disposal Inf...

Страница 11: ...epth 360 mm 3 2 Power requirements Data Value Unit Rated voltage 220 240 V Frequency 50 60 Hz Power 150 W 3 3 Rotational speed Data Value Unit Rotational speed forwards backward 2300 rpm 3 4 Grinding...

Страница 12: ...Manufacturing plate The manufacturing plate is found on the inside of the base of the instrument and contains the following information Item number Type Serial number Rotational speed Voltage Frequenc...

Страница 13: ...ner under the instrument 6 Power switch 7 Power plug 8 Sponge holder with sponge bottom 9 Grinding wheel 10 Sponge top 11 Water valve 12 Sensor only with 2928 10 4 2 Power switch The power switch is u...

Страница 14: ...4 5 Power supply cord 1 Instrument inlet connector instrument 2 Power plug power supply The power supply cord is used to connect the manual grinding wheel to the power supply of the healthcare establ...

Страница 15: ...k Fig 8 1 is used to cool during grinding and is operated using the water valve Fig 8 2 4 8 Contents on delivery Included in the delivery 1 x Hand grinding stone 2928 00 or 2928 10 1 water tank 1 powe...

Страница 16: ...ct the individual components from transport damage corrosion and other damage prior to assembly Therefore do not destroy the carton and unpack the items shortly before assembly Dispose of packaging ma...

Страница 17: ...t expose to aggressive media Protect from sunlight Avoid mechanical shocks Storage temperature 15 to 35 C Relative air humidity max 60 If storing for longer than 3 months regularly check the general c...

Страница 18: ...e to optical measuring instruments Grinding dust and water spray may damage measuring instruments Do not set up the manual grinding wheel near optical measuring instruments Personnel Specialized perso...

Страница 19: ...tach water tank 1 Water tank 2 Attachment surface 3 Sponge with hole 4 Water valve 6 5 Connect the unit to the power supply NOTICE NOTICE The instrument may be damaged by incorrect power supply Incorr...

Страница 20: ...on All work must only be done by skilled personnel Personnel Specialized personnel Protective equipment Safety goggles Safety cap 7 2 Grinding lenses 7 2 1 Check water collection container NOTICE NOTI...

Страница 21: ...below the top Fig 13 Attach water tank 2 Place the water tank Fig 13 1 on the attachment surface Fig 13 2 Ensure that the outlet of the water valve Fig 14 2 is located over the hole in the sponge Fig...

Страница 22: ...ds run direction downwards Fig 15 7 2 4 Switching the instrument on off Personnel n Specialized personnel Switch the power switch to I Fig 16 The grinding wheel rotates Fig 16 Power switch To switch t...

Страница 23: ...th water cooling depends on the material e g silicate glass or CR39 Other ma terials e g polycarbonate or Trivex may only be ground while dry 3 Set the direction of rotation of the grinding wheel Chap...

Страница 24: ...can attack the plastic components Only use mild cleaners that do not contain solvent sand or alcohol 1 Ensure that the instrument is not connected to a power supply pulling the plug out if necessary...

Страница 25: ...1 upwards out from the outer housing 1 Bottom sponge Fig 19 Remove the sponge 3 Rinse the sponge Fig 20 out under running water Fig 20 Bottom sponge If necessary the sponge must be replaced 4 Insert t...

Страница 26: ...device Chapter 7 2 4 Switching the instrument on off on page 22 6 Carefully remove the cleaning dust from the outer housing of the instrument 1 Grinding wheels grit 2 Grinding wheels e g diamond coat...

Страница 27: ...el groove 4 5 Cleaning stone Fig 24 Remove the wheel with the Vbevel groove diagram Continue this process for the 3 areas until the grinding wheel surface is again flat and clean 8 5 Changing the grin...

Страница 28: ...p tacle fuse holder Instrument inlet connector loose Check instrument inlet connector is plugged firmly in Grinding speed is inadequate Grinding wheel blocked up with grinding material Remove material...

Страница 29: ...er accepts old instruments The old instrument can also be broken down into its components and the components sent to recycling Scrap any metal Send plastic elements to be recycled Dispose of remaining...

Страница 30: ...use storage 14 Intended use 7 O Operation 20 Ophthalmic lens material CR39 23 P Packaging 16 Plugs 14 Power supply cord 14 Power switch 13 Protective equipment 10 S Safety General 7 Service 6 Setting...

Страница 31: ...31 Manual Grinding Wheel 2928 00 CE declaration...

Отзывы: