In the case of returns, the general terms and conditions for delivery and payment of
BANDELIN electronic GmbH & Co. KG shall apply.
In addition, the ultrasonic bath must be cleaned and decontaminated (if necessary), see
the upcoming chapter.
6.4.2
Decontamination certifi cate
If the ultrasonic/rinsing bath is sent back to the manufacturer for repairs (with
accessories, if applicable), the form "Certi
fi
cate of Decontamination" must be
fi
lled out
and af
fi
xed to the packaging on the outside, in a visible spot.
If this form has not been
fi
lled out, we reserve the right to refuse receipt of the package
in order to protect our employees.
The form can be downloaded from the Internet as a PDF
fi
le:
www.bandelin.com - Download ...
A sample copy can be found in the appendix.
6.4.3 Replacing fuses
CAUTION!
Repair work may only be carried out by authorised, quali
fi
ed personnel or by the
manufacturer. The manufacturer assumes no liability for unauthorised work performed
on the ultrasonic bath!
The mains plug must be pulled out before opening the ultrasonic bath!
There is a risk of electric shock from live parts in the ultrasonic bath!
Empty the ultrasonic/rinsing bath and disconnect from the mains.
Carefully place it on its back side. Be careful not to bend the mains cable. We
recommend that a pair of suitable wooden slats be placed under the tank while the
replacement process is taking place.
Unscrew the base plate, loosen it and open it.
Unscrew and remove the cover.
The fuses are situated in the fuse switch disconnectors and on the power modules.
After inspection, replace only the defective fuses. The fuse values are listed under
"Technical Data".
Reassemble the ultrasonic bath following the reverse order.
Turn the bath upright and reconnect the mains plug.
The ultrasonic/rinsing bath is once again ready for operation.