In the event of returns, the General Terms and Conditions for Delivery and Payment of
BANDELIN electronic GmbH & Co. KG apply.
In addition, the ultrasonic bath must be cleaned and decontaminated (if necessary), see
the following Section.
6.4.2
Certi
fi
cate of Decontamination
If the ultrasonic bath is sent back to the manufacturer for repairs (with accessories, if
applicable), the form "Certi
fi
cate of Decontamination" must be
fi
lled out and a
ffi
xed to
the packaging on the outside, in a visible spot.
If this form has not been
fi
lled out, we reserve the right to refuse receipt of the package
in order to protect our employees.
The form can be downloaded as a PDF
fi
le from our website:
www.bandelin.com - Downloads ...
A sample copy can be found in the annex.
6.4.3 Replacing fuses
CAUTION!
Repair work may only be carried out by authorised, quali
fi
ed personnel or by the
manufacturer. The manufacturer assumes no liability for unauthorised interventions on
the ultrasonic bath!
The mains plug must be disconnected before opening the bath!
There is a risk of electric shock from live parts in the ultrasonic bath!
Empty the ultrasonic bath and disconnect from the mains.
Carefully place it on its back side. Be careful not to kink the mains cable. We
recommend that a pair of wooden slats be placed under the tank while the
replacement process is taking place.
Unscrew the base plate, loosen it and open it.
The fuses are situated in the fuse switch disconnectors and on the generator
circuit board.
After inspection, replace only the defective fuses. The fuse values are listed in the
"Technical Data".
Reassemble the ultrasonic bath following the reverse order.
Turn the bath upright and reconnect the mains cable.
The ultrasonic bath is ready once again for operation.