1586-006 en/2022-08
25/46
Operation
/5
Switch off the ultrasound before removing the sonication objects.
Do not remove sonication objects by hand. Carefully remove e.g.
the insert basket with the sonication objects and place it on a level
surface.
Rinse sonication objects with clean water.
Do not leave sonication objects in the sonication fluid for too long.
This can damage the objects.
5.12 Empty the oscillating tank
WARNING
Danger of electric shock
– Disconnect the mains plug before lifting the ultrasound bath.
– Do not place the ultrasound bath in a sink for emptying.
– Make sure that no liquid can get into the housing.
CAUTION
Hot sonication fluid and oscillating tank
There is a risk of burns when lifting the ultrasound bath to empty it.
– Allow the ultrasound bath to cool down before lifting it.
Contamination on the bottom of the oscillating tank reduces the
ultrasonic output. Empty and clean the oscillating tank if the soni-
cation fluid is visibly contaminated.
Also, observe the specifications of the manufacturer of the agent
regarding the service life of the sonication fluid.
Replace used sonication fluid completely. Do not freshen the fluid
by topping it up.
Procedure
1. Switch off the ultrasound.
2. Switch off the heating.
3. If you need to move the ultrasound bath to empty it, disconnect
the mains plug.
4. Open the ball valve at the outlet.
5. Rinse the oscillating tank thoroughly.
6. Wipe the ultrasound bath dry with a soft cloth.