The frequency scan is completed once
fi
ne-tuning has been successful, and the display switches
back to stand-by mode.
Otherwise, an error message such as Error 012 appears again.
6.3
Error analysis
Errors may arise
• on the plug connections
• on the ultrasonic converter
• on the probes
• on the ultrasonic generator
The device is robustly constructed and designed for a high level of reliability. Nevertheless,
the possibility of a malfunction due to a defective component can never be fully discounted.
Mechanical defects of the connection for ultrasonic converter, the plug connectors, the ultrasonic
converter, etc., are possible as a result of frequent use or even incorrect handling, e.g. by dropping
them.
Critical faults are recognised by the device and signalled by a red LED and an intermittent signal
tone (3 times), and displayed with an error number, see next page.
Error no.
Possible cause
Remedial action
Converter:
No USC
No ultrasonic converter (USC) connected
to the ultrasound generator (GM)
Connect the USC to the horn/probe on the GM. The
GM should then automatically recognize which USC
has been plugged in.
002
Frequency setting not possible
Perform frequency scan. If the error occurs again,
contact BANDELIN.
003
No power output, amplitude setting not
possible
Error correction should only be carried out by
BANDELIN. Send in the complete ultrasonic
homogeniser.
004
Mains voltage interrupted
Check mains connection, check fuse
010
Frequency synchronization
malfunctioning, frequency display is not
currently possible
- Check whether the probe has been mounted
correctly = su
ffi
ciently tight
- Check for damage:
• Horn/probe contact surfaces
• Horn/probe
threads
• Probe radiation surface
If needed, replace the USC/probe and then check its
function. Use the “probe check” or “scan frequency”
functions for this purpose. After successfully
completing the frequency scan, do not restart
operation until initialization is complete. In the event
of failure, contact BANDELIN.
011
No return signal from USC, USC or probe
defective
012
Error in frequency scan, no resonance
frequency found, probe defective
014
Permissible internal operating
temperature exceeded, ultrasonic
homogenizer switches o
ff
Allow the ultrasonic homogeniser to cool down
for at least 10 minutes, if necessary check with a
temperature sensor that the operating temperature
of the GM has dropped to < 50 °C.