Ultrasonic converter
Standard/booster horn
Vigorously turn sickle spanner
HS 40/42 long, in the opposite
direction
30 / 46
51376-010 GB/2020-10
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
4.4
Dismounting and mounting the standard/booster horns
Standard/booster horns
• are screwed onto the respective ultrasonic converter,
• are made of high-strength titanium alloy,
• transmit the oscillations to the probe,
• enhance the amplitude thanks to their geometry.
Standard/booster horn SH 100 G is
fi
rmly mounted onto the UW 100 and SH 200 G onto UW 200
at the time of delivery.
Before mounting the standard/booster horns, the ultrasonic generator must
be turned o
ff
and the ultrasonic converter must be disconnected from the
ultrasonic generator.
Video available under:
https://www.youtube.com/watch?v=NXbGc6nAb5c
Dismounting
Release
the
fi
rmly-mounted standard/booster horn SH from the ultrasonic converter UW. To do
so, lay the ultrasonic converter on a
fi
rm base (non-slip if possible).
Place one sickle spanner on the ultrasonic converter and the other sickle spanner on the
standard/booster horn. To dismount, press the
fi
rst sickle spanner against the
fi
rm base, press
the other sickle spanner
fi
rmly in the opposite direction.
One
sickle spanner is part of the s cope of delivery of one standard/booster horn.