Conducting the foil test
1. Fill oscillating tank to the filling level mark with water
and an appropriate ultrasonic agent, in the concentration
specified by the manufacturer.
2. Degas the liquid (see user instructions)
3. Stretch aluminium foil (household foil, 10
v
m to 25
v
m
thick) over the foil testing frame.
Depending on the tank size, it is possible that the frame
will protrude outside the tank.
It is sufficient to cover the submerged portion of the
frame with foil.
Frame Ø 3 mm
Aluminium foil, folded 10 mm
along the edge
4. With the ultrasound switched off,
position or fix the foil-wrapped
frame at an angle across the middle
of the oscillating tank (see video).
D
ev
ic
e, f
ro
n
t
Devic
e,
rear
939aa GB/2020-07
5. Switch on the ultrasound and sonicate the foil for at least
one minute until visible perforations or holes are produced.
With sturdier foils (thicker or coated ones), the sonication
time may be up to 3 minutes.
6. Switch off the ultrasound, take the foil out
and let it dry.
7. The foil must be perforated,
see photo. Otherwise, we
recommend having the
device checked by the service
department at BANDELIN
electronic GmbH & Co. KG.
8. Archiving of foil with test date
and serial number of the ultrasonic bath. The foil test
document template can also be completed and archived.
9. After the test, the oscillating tank must be thoroughly
rinsed out to remove any loose foil particles.
Type
Order no.
for
FT 1
3190
DT 31/H, DT 52/H
RK 31/H, RK 52/H
FT 4
3074
DL 102 H, DL 255 H, DT 100/H, DT 102H/H-RC,
DT 103, DT 106, DT 255/H/H-RC,
RK 100/H, RK 102 H, RK 103, RK 106, RK 255/H
FT 6
3222
DL 156 BH, DT 156/BH,
RK 156/BH
FT 14
3084
DL 510 H, DL 512 H, DL 514 BH, DT 510/H/H-RC,
DT 512 H, DT 514H/BH/BH-RC, DT 510 F, RK 510/H,
RK 512 H, RK 514/H/BH, ZE 514/...DT
FT 36
3673
DT 1028 F, ZE 1031/1032/...DT
FT 37
3674
DT 1058 M, ZE 1058/1059/...DT
FT 38
3672
MC 1001/E
FT 40
3094
DL 1028 H, DT 1028/H/CH, RK 170 H,
RK 1028/H/C/CH, RK 1040
FT 42
3224
TRISON (TE 3000)
FT 45
3204
DT 1050 CH,
RK 1050/CH
2 / 2
Suitable foil testing frames can be ordered from
BANDELIN electronic GmbH & Co. KG.
The foil testing frames are suitable for a wide range of tank
dimensions. Aluminium foil is also required for conducting
the test, but this is not included in the delivery.
BANDELIN electronic GmbH & Co. KG
Heinrichstraße 3–4
12207 Berlin
Deutschland
www.bandelin.com
[email protected]
+ 49 30 768 80 - 0
+ 49 30 773 46 99
Zertifiziert nach
ISO 9001