background image

Page 18

194 M1 01 F

12-13-2013

ELECTRODE CALIBRATION

ENTER

ENTER

ENTER

ENTER

ENTER

VALID

CAUTION - Temperature compensation: 
set up the instrument on MANUAL mode at 20°C

During  calibration  in  mode  “MODIFICATION”,  measurement
value is on stand-by on the last reading value, the regulation is on
stand-by. A delay after ending the calibration allows a restart of
the complete system before to automatically restart the regulation
mode. 

Choose 

NO

, the sensor is a non adjusting zero required 

NO

will send to the display “

SLOPE

Choose 

YES

to proceed to a sensor calibration 

NO

will send to the display “

DELAY

” 

The  calibration  measuring  value  is  locked  when  the  submenu
‘’STANDARD’’ is displayed. Proceed at the same time to a test to
know the standard value according to the sensor in use.

Enter the result value

(For good calibration, value has to be as closed as the sensor
full range)

Sensor gain is displayed.

CAUTION: 

If the slope value is >200% or <50%,

do not valid this calibration

step.  Check  the  dirtiness  of  the  sensor;  do  a  maintenance  on  it
and proceed to a second calibration.

If the slope value is between 50 and 200%, 

proceed to a second

test to confirm the previous test 

Set  up  the  time  during  the  measuring  and  status  regulation

previous  to  the  calibration,  will  be  displayed  after  saving  the
calibration. 

CAUTION - Temperature compensation: if a PT100 probe is
connected, reset the temperature compensation in "AUTO"

STANDARD. ZERO 

YES/NO

STANDARD. SLOPE 

YES/NO

STANDARD  

00.00

ppm

SLOPE    xxx,x  %

DELAY  0015 Sec

SAVING ?

MENU

+

Next menu:

FORCED RELAY

Содержание BAMOPHOX 194

Страница 1: ...MES 194 01 1 Amperometric signal monitor BAMOPHOX 194 13 12 2013 194 M1 01 F INSTRUCTIONS MANUAL BAMOPHOX 194 Amperometric sensors signal monitor...

Страница 2: ...ONT PANEL 7 SCROLLING MENU 8 ABOUT BAMOPHOX 9 CONSULTING MODIFYING 9 TESTING RELAY OUTPUTS 9 SENSOR CONFIGURATION 10 ADJUSTING THRESHOLD 1 11 ADJUSTING THRESHOLD 2 11 ADJUSTING THRESHOLD 3 12 ON OFF R...

Страница 3: ...C depending of switching frequency ambient conditions accuracy Mechanical life time minimum 5 x106 operations 180 commutation min Electrical life time minimum 2 x105 20 comm min 3 A 125 V AC 3 A 30 V...

Страница 4: ...M1 01 F 12 13 2013 WALL MOUNTINg INSTRUMENT 3 WIRINg OPTION LOGGER RS 422 OPTION LOGGER access by upper cover RS 422 PANEL MOUNTINg INSTRUMENT 29 30 31 36 V V 0V E Signal input Signal input 29 30 31 3...

Страница 5: ...ntact Measurement Temperature S3 threshold NO contact Measurement Temperature External sensor NOT USED Ground N Main power supply N Neutral L Line L V V 0V E 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17...

Страница 6: ...SENSOR CONNECTION Load resistor supplied with the BAMOPHOX 194 4 20 mA sensor 0 0 to 2000 mV sensor TERMINAL FOR SENSOR CONNECTION Flow rate sensor 194 832 NPN type normally open Brown Blue Black outp...

Страница 7: ...his key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input in...

Страница 8: ...ADJUST ALARM 1 ADJUST ALARM 2 ADJUST ALARM 3 RELAY REGULATION REGUL PID OUTPUT mA MES OUTPUT mA TEMP TEMPERATURE SENSOR CALIBRATION FORCED RELAY ADJUST ALARM CLOCK RECORDING TIME LANGUAGE MENU MENU M...

Страница 9: ...the configuration thresholds regulation analog outputs When PID regulation is activated the display shows the PID FORCED PID 00 00 ppm 00 00 One digit is flashing Modify the value Immediately the inst...

Страница 10: ...D regulation with the 4 20 mA signal There is a specificity of alarm 3 in Bamophox 194 It can be setup for an external sensor Signal is coming from an inductive flow rate sensor type NPN This sensor...

Страница 11: ...e contact S1 SAVINg ADJUST ALARM 2 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against measured value TEMP Threshold against temperature measured value ALARM 2 LOW HIgH HIgH Contact closes w...

Страница 12: ...he limit LOW Contact closes when the value goes under the limit ON 00 00 ppm To close the contact S3 at this value OFF 00 00 ppm To open the contact S3 at this value DELAY UP ONI OFF Delay or no delay...

Страница 13: ...est value for LOW PB 00 00 ppm proportional bandwith Highest value H IGH DZ 00 00 ppm for dead zone Lowest value LOW DZ 00 00 ppm for dead zone SAVING ENTER ENTER ENTER ENTER ENTER ENTER ENTER ENTER E...

Страница 14: ...tion contact S1 is closed Inside the dead zone between 1 4 1 9 ppm S1 and S2 are OFF no injection at all contacts S1 and S2 are opened If the measurement value is between the dead zone and the highest...

Страница 15: ...such as m Cde t delay of response T Time corresponding to the same variation in of measurement m and the analog output Cde m Cde This value may be found out on the slope Modify the PID parameters as f...

Страница 16: ...Page 16 194 M1 01 F 12 13 2013 ppm measurement...

Страница 17: ...lue corresponding to 00 00 or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING ENTER ENTER ENTER ENTER VALID TEMPERATURE AUTO A Pt 100 probe should be connected MEASURE AUTO MANUAL MANUAL Without Pt...

Страница 18: ...ARD is displayed Proceed at the same time to a test to know the standard value according to the sensor in use Enter the result value For good calibration value has to be as closed as the sensor full r...

Страница 19: ...g a malfunction on S1 and S2 contacts an overtime contact could be set up WITH WHITHOUT ALARM TMAX S1 0000 Sec Over timing on S1 closed contact maximum time for active relay TMAX S2 0000 Sec Over timi...

Отзывы: