Page 18
194 M1 01 F
12-13-2013
ELECTRODE CALIBRATION
▲
▲
▲
▲
▲
▲
ENTER
ENTER
ENTER
ENTER
ENTER
VALID
CAUTION - Temperature compensation:
set up the instrument on MANUAL mode at 20°C
During calibration in mode “MODIFICATION”, measurement
value is on stand-by on the last reading value, the regulation is on
stand-by. A delay after ending the calibration allows a restart of
the complete system before to automatically restart the regulation
mode.
Choose
NO
, the sensor is a non adjusting zero required
NO
will send to the display “
SLOPE
”
Choose
YES
to proceed to a sensor calibration
NO
will send to the display “
DELAY
”
The calibration measuring value is locked when the submenu
‘’STANDARD’’ is displayed. Proceed at the same time to a test to
know the standard value according to the sensor in use.
Enter the result value
(For good calibration, value has to be as closed as the sensor
full range)
Sensor gain is displayed.
CAUTION:
If the slope value is >200% or <50%,
do not valid this calibration
step. Check the dirtiness of the sensor; do a maintenance on it
and proceed to a second calibration.
If the slope value is between 50 and 200%,
proceed to a second
test to confirm the previous test
Set up the time during the measuring and status regulation
previous to the calibration, will be displayed after saving the
calibration.
CAUTION - Temperature compensation: if a PT100 probe is
connected, reset the temperature compensation in "AUTO"
STANDARD. ZERO
YES/NO
STANDARD. SLOPE
YES/NO
STANDARD
00.00
ppm
SLOPE xxx,x %
DELAY 0015 Sec
SAVING ?
MENU
+
Next menu:
FORCED RELAY
Содержание BAMOPHOX 194
Страница 16: ...Page 16 194 M1 01 F 12 13 2013 ppm measurement...