background image

Page

3

436 M1 01 C

1.  TECHNICAL FEATURES

Displayed parameters: 

Measurement values of Turbidity - Configuration Menu - Temperature value

Display: 

Back lighted - 2 lines of 16 alphanumerical characters ; 9,2 mm high

Indication: 

LED alarms status

Configuration: 

8 push buttons keyboard on front face - Keyword protected

Units: 

NTU, FAU, mg/L, g/L, FNU 

Scale

Set up from 0.001 to 9999, adjustable decimal point 

Turbidity measurement: 

input 0/4-20 mA

Temperature: 

input for a 3 wires Pt 100 Ω /0°C

Relay outputs: 

4 closing contacts (Silver alloy), voltage free
3 programmable independent thresholds - with adjustable hysteresis 0…100% - and 
adjustable timer from 0 to 9999 sec
1 Output relay (S4) as a common alarm

Contact: 

Switching capacity 250 V / 3A AC, 30 V / 3A DC, minimum 100 mA / 5 V DC

Measurement output: 

0/4-20 mA (maxi 600 Ω) proportional to measurement, galvanic insulated

Temperature output: 

0/4-20 mA (max 600 Ω), scaling 0…100°C, galvanic insulated

Main power supply: 

230 V AC / 50-60 Hz [other on request] - Consumption 10 VA

Models: 

Panel mounting, front IP65, 72 x 144 mm, connections on screw terminal IP40
Idem DIN Rail mounting, only for blind monitor, IP40
Wall mounting, IP65, cable glands, connections on screw terminal

OPTION

(RS 422 + Logger) 

Communication: 

RS422 output, J-BUS link, binary slave mode, 2400 to 9600 bauds

Data Logger: 

Cycle average measurement record, with a programmable period,
150000 records on MMC (multi media card) / External driver necessary for reading

CE label

Labelized CE in conformity with 73/23/CEE low power and electromagnetic compatibilty 89/336/CEE

2.  CAUTION

- The instrument may not be subject to vibrations and should be protected against direct sun shining. 

The ambient temperature  should be between 0 and 50°C. 

- Wiring have to be done by a specialist. 
- Any error on electrical connection may cancel terms of warranty. 
- Before to switch on your instrument, please check that the main power supply corresponds to the device requirement. 

3.  DIMENSIONS

Extension terminal:

identical to the panel
or wall mounting

PANEL

Wall mounting instrument

Panel mounting instrument

Содержание 436-02

Страница 1: ...bamo fr Fax 33 01 34 10 16 05 E mail info bamo fr Turbidity monitor and controller BAMOPHOX 436 MES 436 01 1 INSTRUCTION MANUAL INSTRUCTION MANUAL B B B BA A A AM M M MO O O OP P P PH H H HO O O OX X...

Страница 2: ...ontents 1 TECHNICAL SPECIFICATIONS Page 3 2 CAUTION 3 3 DIMENSIONS 3 4 WIRING 4 5 FRONT PANEL 7 SCROLLING MENU 8 ABOUT BAMOPHOX 8 CONSULTING MODIFYING 8 MEASUREMENT PARAMETERS 9 ADJUSTING THRESHOLD 1...

Страница 3: ...supply 230 V AC 50 60 Hz other on request Consumption 10 VA Models Panel mounting front IP65 72 x 144 mm connections on screw terminal IP40 Idem DIN Rail mounting only for blind monitor IP40 Wall mou...

Страница 4: ...Page 4 436 M1 01 C 4 WIRING WALL MOUNTING OPTION LOGGER RS 422 OPTION LOGGER access by the upper cover RS 422 PANEL MOUNTING Measurement input...

Страница 5: ...ld NO contact Measurement Temperature S3 threshold NO contact Measurement Temperature NOT USED Ground N Main power supply N Neutral L Line L NC mA Turbidity transmitter mA 1 2 3 4 5 6 7 8 9 10 11 12 1...

Страница 6: ...blind BAMOPHOX B A S4 contact NO alarm on temperature or over range S2 threshold NO contact Measurement Temperature S1 threshold NO contact Measurement Temperature S3 threshold NO contact Measurement...

Страница 7: ...key all parameters are saved previous data programmation will be overwrited If you are not sure of any modification do not press the VALID key To change parameters of data capture Num ric input increa...

Страница 8: ...MENU MENU MENU MENU MENU MENU MENU MENU MENU ABOUT Bamophox ABOUT Bamophox VERSION 2 04 SERIAL N 20451 05 CONSULTATION MODIFICATION CONSULTATION MODIFICATION CODE 0000 CODE 5105 Last 4 digits of seria...

Страница 9: ...NTU Decimal point selection SCALE MINI 00 00 NTU Minimal range value MAXI 00 00 NTU Maximal range value TRANSMITTER MINI 04 00 mA Minimal calculated value to display MAXI 20 00 mA Maximal calculated v...

Страница 10: ...1 SAVING ADJUST ALARM 2 ADJUST ALARM 3 page 11 ALARM 2 ON OFF ALARM 2 MEASURE TEMP MEASURE Threshold against pH mV measured value TEMP Threshold against temperature measured value ALARM 2 LOW HIGH HIG...

Страница 11: ...DELAY UP ONI OFF Delay or no delay before to close the contact S3 TIME 0000 SEC Delay time to close the contact S3 DELAYDOWN ON OFF Delay or no delay before to open the contact S3 TIME 0000 SEC Delay...

Страница 12: ...rresponding to 00 00 mA or 04 00 mA OUTPUT 4 20 mA 0 20 mA Output type SAVING L L L L L ENTER ENTER ENTER ENTER VALID MENU FORCED RELAY LANGUAGE ALARM 1 ON OFF ALARM 2 ON OFF Diagnostic mode to test t...

Отзывы: