Page 4
2. SPECIFICATION
2.1 GENERAL SPECIFICATION
Accuracy: 0.05%rdg, 0.02%rdg, 0.01%rdg, 0.005%rdg
Piston and cylinder material: Tungsten carbide
Mass material: 304SS (Forged 316 stainless steel is optional)
Tester base and pump material: Stainless Steel
Medium: Dry and clean air, or nitrogen
Pressure connector: Metric, (NPT and BSP is optional)
Gravity: Standard 9.80665 m/s², or customize local gravity.
Tester dimension: 500mm(H) x 650mm(L) x 500mm(D) / Pump dimension: 225mm(H) x 455mm(L) x 380mm(D)
Tester weight: Appro.25kg / Pump weight: Appro.15kg
Recalibration: Every 2-5 years
2.2 PRESSURE RANGE SELECTABLE SHEET
P/N
Pressure
Accuracy
Area
Kn. factor
Mass Set
KYQZ100
(-100~100) kPa
0.05, 0.02, 0.01, 0.005
10cm
2
10kPa / kg
10kg + 10kg
KYQZ160
(-100 ~ 160) kPa
0.05, 0.02, 0.01, 0.005
5 cm
2
20kPa / kg
5kg + 8kg
KYQZ250
(-100 ~ 250) kPa
0.05, 0.02, 0.01, 0.005
5 cm
2
20kPa / kg
5kg + 12.5kg
KYQZ400
(-100 ~ 400) kPa
0.05, 0.02, 0.01, 0.005
5 cm
2
20kPa / kg
5kg + 20kg
KYQZ600
(-100 ~ 600) kPa
0.05, 0.02, 0.01, 0.005
2 cm
2
20kPa / kg
5kg + 30kg
KYQZ1M
(-100 ~ 1000) kPa
0.05, 0.02, 0.01
2 cm
2
20kPa / kg
5kg + 50kg
Содержание BL506