Se lec tion  of  a  pa ra me ter

se lect the pa ra me ter by pres sing  “+” -bu tton oder “-” -bu tton

Con firm the pa ra me ter by pres sing the “P” -bu tton.

The dis play shows the last pro gram med va lue of the se lec ted pa ra me ter

Change  and  con firm  a  se lec ted  pa ra me ter

Change the pa ra me ter by pres sing the “+” -bu tton or the “-” -bu tton

• Se lect the dis play de ca de to be chan ged by pres sing the “*” -bu tton 

Con firm the pa ra me ter by pres sing the “P” -bu tton

The dis play shows the pro gram ming level and the num ber of the pa ra me ter
e.g.: “0-04" => pa ra me ter 4

7.1.

Programming level for configuration 0-00

Param.

Description

Range

Default

Value

0-00

Input range
  0 -> Voltage 0 bis 10 V
  1 -> Current 0 bis 20 mA
  2 -> Current 4 bis 20 mA

0 .. 2

0

0-01

Display value of  min. input signal

-999 .. +9999

0

0-02

Display value of max. input signal

-999 .. +9999

+1000

0-03

Decimal point
0 ->

XXXX

1 ->

XXX.X

2 ->

XX.XX

3 ->

X.XXX

0 .. 3

1

0-04

Tare on = 1, off = 0

0 .. 1

0

0End

Leaving programming level 0-00

11

7. Programming

Содержание BDD-UM 3023

Страница 1: ...english User s Guide BDD UM 3023...

Страница 2: ...efective material and workmanship for a period of two 2 years from date of delivery If it is necessary to return the product to us the sender is responsible for shipping charges freight insurance and...

Страница 3: ...ns 6 4 1 General Instructions 6 4 2 Hints against noisy enviroment 7 4 3 Connection and pin assignment 7 4 4 Connection of input signals 8 4 4 1 Input voltage 0 10 V 8 4 4 2 Input current 0 20 mA 4 20...

Страница 4: ...7 1 1 Scaling the display range 12 7 2 Tare function 13 7 3 Display test 13 7 4 Programming quick reference 13 8 Technical specifications 14 3...

Страница 5: ...lity to customer or any other person or entity with respect to any liablity loss or damage caused or alleged to be caused directly or indirectly by equipment or software sold or furnished by us Read t...

Страница 6: ...ion 3 Mounting 3 1 Place of operation Attention must be payed to the protection against humidity dust and high tempe ratures at the place of operation 3 2 Panel mounting 3 2 1 Panel for switch board I...

Страница 7: ...tages applied Cords must be provided with sleeves Attention must be paid that the power supply voltage applied will agree with voltage noticed at the Type plate It is important to ensure a careful gro...

Страница 8: ...advisable To use shielded cables The wiring of shields and ground 0V should be star shaped The distance to interference sources should be as long as possi ble If necessary protective screen or metal e...

Страница 9: ...3 Signal ground 6 Supply voltage 4 4 Connection of input signals 4 4 1 Input voltage 0 10 V 4 4 2 Input current 0 20 mA 4 20 mA 4 5 Connection of power supply voltage 8 4 Electrical connections 3 0 V...

Страница 10: ...onfiguration by programming Attention When the instrument is built in a machine and the custo mer wants to change the configuration attention must be paid that no damage will occur to the machine 6 Pr...

Страница 11: ...y the buttons on the front Push button Function P button Selection of parameter button Increase of number of parameter parameter button Decrease of number of parameter parameter button select the disp...

Страница 12: ...on Confirm the parameter by pressing the P button The display shows the programming level and the number of the parameter e g 0 04 parameter 4 7 1 Programming level for configuration 0 00 Param Descri...

Страница 13: ...input current 1mA or 3mA the display range exceeds the possible minimum 999 the parameter P0 03 has the value3 decimal point after the first decade X XXX in this case is the display of negative value...

Страница 14: ...0 04 had to be set to 1 Erase the tare value push the button for ca 3 seconds After this reset the untared value is dis played 7 3 Display test Activating the display test will drive all segments of...

Страница 15: ...n at negative Values Operation keyboard design front membrane with push buttons Power supply 24 V 20 DC isolated power consumption max 65 mA Case 96 x 48 x 60 mm depth 72 mm incl screw terminal protec...

Страница 16: ...41042 Phone 859 727 2200 Toll free 1 800 543 8390 Fax 859 727 4823 technicalsupport balluff com CN Service Center China Balluff Shanghai trading Co ltd Room 1006 Pujian Rd 145 Shanghai 200127 P R Chin...

Отзывы: