22
Table 5
Table 5 has been excluded from this manual as it does not pertain to this particular
product.
Table 6
Recommended separation distances between portable and mobile RF
communications equipment and the ME EQUIPMENT or ME SYSTEM
– for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
RECOMMENDED SEPARATION DISTANCES BETWEEN
PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AT THE MODEL # 10500
The Balance Living
TM
Precision LIFE TRACK body analysis scale is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Balance Living
TM
Precision LIFE TRACK body analysis scale can
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Balance Living
TM
Precision LIFE TRACK body analysis scale as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
N/A
0.117
0.233
0.1
N/A
0.369
0.738
1
N/A
1.167
2.333
10
N/A
3.690
7.377
100
N/A
11.67
23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be
determined using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.