Table 2
Guidance and MANUFACTURER’S declaration –
Electromagnetic IMMUNITY – for all ME EQUIPMENT and ME SYSTEMS
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Balance Living
TM
Precision LIFE TRACK body analysis scale is intended for use in the electromagnetic environment specified
below. The user of the Balance Living
TM
Precision LIFE TRACK body analysis scale should assure that it is used in such an environment.
IMMUNITY Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment – Guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast transient / burst
IEC 61000-4-4
±2 kV for power supply lines
Not applicable
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
Not applicable
Voltage dips, short interruptions
and voltage variations on
power supply input lines IEC
61000-4-11
<5% U
T
(﹥95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(﹥95% dip in U
T
) for 5 sec
Not applicable
Power frequency
(50 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Table 3
Table 3 has been excluded from this manual as it does not pertain to this particular
product.