Technical Specifications
12/3/2009 | 71-022 V6
D30R
80
HiPot measurement accuracy – HiPot voltage
Range Resolution
1250V/Div
+/- 5% from 1250V – 7000V
2500V/Div
+/- 5% from 2500V – 10000V
5000V/Div
+/- 5% from 5000V – 20000V
7500V/Div
+/- 5% from 7500 – 30000V
Testing Accuracy - HiPot Measurements
Range Approximate
maximum
measurable
current
Resolution Full
scale
accuracy
100
μ
A/Div
Range
900
μ
A +/- 5%
or 45
μ
A
+/-5% from
90
μ
A-900
μ
A
+/-5%
10
μ
A/Div
Range
90
μ
A +/- 4.5
μ
A
+/-5% from
9
μ
A - 90
μ
A
+/-5%
1
μ
A/Div
Range
9
μ
A +/-.45
μ
A +/-5%
from
.9
μ
A - 9
μ
A
+/-5%
Voltage measurement accuracy – Surge (D30R)
Range Resolution
1250V/Div
+/- 12% from 500V – 1960 V
2500V/Div
+/- 12% from 500V – 3930 V
5000V/Div
+/- 12% from 1000V – 7870 V
7500V/Div
+/- 12% from 2000V – 11900 V
Calibration information
Please contact Baker Instrument Company for current calibration information. Contact the
service department at (970) 282-1200, or (800) 752-8272.
Содержание D30R
Страница 6: ...Table of Contents 12 3 2009 71 022 V6 D30R 6...
Страница 12: ...Safety precautions CE compliance 12 3 2009 71 022 V6 D30R 12...
Страница 24: ...Coil Resistance Testing 12 3 2009 71 022 V6 D30R 24...
Страница 28: ...Principles of High Voltage DC Testing 12 3 2009 71 022 V6 D30R 28...
Страница 38: ...High Voltage DC Testing 12 3 2009 71 022 V6 D30R 38...
Страница 50: ...Surge Testing 12 3 2009 71 022 V6 D30R 50...
Страница 70: ...Data Recording Retrieving 12 3 2009 71 022 V6 D30R 70...