Safety precautions, CE compliance
12/3/2009 | 71-022 V6
D30R
10
D30R
Power cord
Operation manual
Pollution Degree II
(From IEC 61010-1 3.6.6.2) Only non-conductive pollution occurs. However, temporary
conductivity caused by condensation is expected.
Power requirements
Using the provided AC power cord, connect the unit to a grounded AC power source. The
unit’s power requirements are 100-240VAC, 50-60 Hz, 2 amps AC maximum current draw.
The unit is fused using 2.5A fast blow fuses. Replace fuses with like type and rating.
Environmental conditions
1)
The unit has been tested for use up to 2000 m.
2)
The tester should only be operated in temperatures ranging from 41 to 104
degrees Fahrenheit (5° C to 40° C).
3)
This unit is for use at a maximum relative humidity of 80% for temperatures up to
31 °C decreasing linearly to 50% relative humidity at 40°C. This unit is intended for
Installation Category II in a Pollution Degree II environment.