Figure 24. Pushbutton
6.
Press the pushbutton into the pump housing.
At best, take the pump in both hands and grasp the cartridge housing with
one hand and press the pushbutton with the other palm of your hand.
The CO
2
gas flows into the device.
7.
The pressure must be constantly checked on the manometer, holding the
pushbutton for a few seconds at the desired pressure so that the desired
pressure also prevails in the camera.
It must not exceed the maximum pressure of the device.
The pushbutton is limited by a safety ring. This must not be removed or
manipulated.
Maximum achievable pressure (pushbutton pressed in until stop): less than 2.5
bar (36 psi).
When the filling process is finished, the push button is released. The pressure in
the pump is reduced. The pressure display decreases.
8.
The valve adapter can now be removed.
Figure 25. Screw in valve seal