Chapter 4. Ethernet Only Communications
44
Communications Options User’s Guide
4.2.5 Changing IP Parameters
Example:
Change the DHCP-assigned IP address to static address 192.168.2.207 and disable DHCP for the unit with IP assigned
address 3.112.162.129 (see Figure 24).
1.
Under Device Task select
Open web interface.
2.
Enter the Username and Password. The factory defaults are
root and dbps.
Note:
The username and password are case sensitive.
Panemetrics recommends changing the default password and disabling unused ports and services.
3.
Click on
Login
.
4.
Select
Configuration / Network.
5.
Select
Use the following IP address: and enter IP address
192.168.2.207
.
6.
Click on
Apply
.
Note:
Changes will require a reboot to take effect.
Select
Administration / Reboot, then wait for the reboot to complete.
Figure 24: Changing IP Parameters
4.2.6 Changing User’s Information
To change the user name and/or password:
1.
Under Device Task select
Open web interface.
2.
Enter the Username and Password. The factory defaults are
root and dbps.
Содержание Panametrics XMT686i
Страница 1: ...panametrics com Flow 916 115 F JUL 2021 Communications Options User s Guide...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 90: ...Chapter 8 HART Communications 80 Communications Options User s Guide no content intended for this page...
Страница 98: ...Warranty 88 Communications Options User s Guide no content intended for this page...
Страница 99: ......