AquaTrans™ AT600 User’s manual
135
Chapter 6. Communication
6.3 Additional device status
Command 48 returns 4 bytes of data, with
Table 185: HART additional device status
HART additional device status
Class
Device status
bits set
Byte
Bit
Error description
0
0
Amplitude error
Error
4, 7
1
Low signal
Error
4, 7
2
Sound speed error
Error
4, 7
3
Velocity range
Error
4, 7
4
Signal quality
Error
4, 7
5
Cycle skip
Error
4, 7
6
Reserve
7
Reserve
1
0
Reserve
1
Reserve
2
Reserve
3
Reserve
4
Reserve
5
Reserve
6
Reserve
7
Reserve
2
0
FPGA error;
4, 7
1
Setting files CRC error;
4, 7
2
Flash error
4, 7
3
KEY/LED error
4, 7
4
I/O error
4, 7
5
Display error
4, 7
6
RTC error
4, 7
7
Reserve
3
0
In configure mode;
4, 0
1
Not calibrated;
4, 0
2
Reserve
3
Reserve
4
Reserve
5
Reserve
6
Reserve
7
Reserve
Содержание Panametrics AquaTrans AT600
Страница 1: ...panametrics com BH001C11 Rev F AquaTrans AT600 User s manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii AquaTrans AT600 User s manual no content intended for this page...
Страница 20: ...Chapter 1 General installation instructions 10 AquaTrans AT600 User s manual no content intended for this page...
Страница 36: ...Chapter 3 Wiring the AT600 electronics 26 AquaTrans AT600 User s manual no content intended for this page...
Страница 84: ...Chapter 4 Initial setup and programming 74 AquaTrans AT600 User s manual no content intended for this page...
Страница 150: ...Chapter 6 Communication 140 AquaTrans AT600 User s manual no content intended for this page...
Страница 158: ...Appendix B Specifications 148 AquaTrans AT600 User s Manual no content intended for this page...
Страница 162: ...Appendix C Specifications 152 AquaTrans AT600 User s Manual no content intended for this page...
Страница 164: ...Appendix D Menu maps 154 AquaTrans AT600 User s Manual D 2 The main menu Figure 42 The main menu...
Страница 172: ...Warranty 162 AquaTrans AT600 User s Manual...
Страница 173: ......