The display can show either: POD = Off or POD = current value (QFE pressure 

reading). The POD RESET key (F2) sets the current POD correction to Off.

2.1.1  Method (Accessible Switch Contacts)

1. Connect the instrument to the static system and connect the altitude switch as shown. 

Ensure that the pitot system is vented to atmosphere before starting the test:

Note: 

Contacts must be potential free.

Figure

 

3: Altitude Switch Test Connections

2. Select negative pressure and close the release valve.
3. Close the manual let-down valve.
4. Press the RUN key (F1) and operate the hand-pump, continue until the switch operates:

mbar

POD = Off

POD Correction

mbar

POD Correction

POD = Off

WARNING

VENT SYSTEM

BEFORE SETTING POD

WARNING

VENT SYSTEM

BEFORE SETTING POD

WARNING

VENT SYSTEM

BEFORE SETTING POD

CHANGE POD 

VALUE

POD RESET

CHANGE POD 

VALUE

POD RESET

CHANGE POD 

VALUE

POD RESET

ENTER VALUE

mbar

POD Correction

POD = 1002.85

ft

F2

F1

S

Applied Static System Pressure

Vent Pitot System (Atmospheric Pressure)

Altitude Switch

Select
Negative (-)
Pressure

SWITCH TEST

CONTACT STATE 

POD = 1008.63 

ALTITUDE 

RUN

ft

ft

ft

SWITCH TEST

CONTACT STATE

POD = Off

ALTITUDE

RUN

MANUAL 

OPERATION

STOP

MANUAL 

OPERATION

STOP

POD = Off

ALTITUDE

POD = Off

ALTITUDE

CONTACT STATE
CLOSED AT 

5550 

ft

OPENED AT 

5325 

ft

HYSTERESIS 225 

ft

SWITCH TEST

SWITCH TEST

CONTACT STATE
CLOSED AT 

5550 

ft

Содержание Druck DPI 610A

Страница 1: ...Druck com DPI 610A Aeronautical Portable Pressure Calibrator Instruction Manual...

Страница 2: ......

Страница 3: ...acturer s service departments Druck com For technical advice contact the manufacturer Symbols WARNING Before operating this instrument read the safety instructions Do not apply pressures greater than...

Страница 4: ...18 ft 30 000 ft 36 ft Precision includes linearity hysteresis and repeatability 12 months measurement stability Temperature Effects 0 004 of reading C averaged over 10 to 40 C w r t 20 C 0 002 of read...

Страница 5: ...g absolute version of the DPI 610 provides manual generation of pressure and vacuum for testing of aircraft Pitot and Static systems It has a special pneumatic manifold and output port assembly the re...

Страница 6: ...ator Controls 1 Output port AN4 to pitot or static system 2 Manual let down valve 3 Electrical switch inputs 4 Release valve releases system pressure vacuum at a controlled rate 5 Vent port 6 Selects...

Страница 7: ...imum measurement ratings of the standard instrument that should not be exceeded Pressure over range 120 of FS is indicated by a flashing pressure display Voltage or current over range 110 of max cause...

Страница 8: ...n POD by selecting POD ADJUST as follows Apply QFE pressure reading as POD correction as follows WARNING Observe safety precautions stated in local orders and the aircraft or equipment servicing proce...

Страница 9: ...RUN key F1 and operate the hand pump continue until the switch operates mbar POD Off POD Correction mbar POD Correction POD Off WARNING VENT SYSTEM BEFORE SETTING POD WARNING VENT SYSTEM BEFORE SETTIN...

Страница 10: ...m is vented to atmosphere before starting the test 2 Select negative pressure and close the release valve 3 Close the manual let down valve 4 Press the RUN key F1 and operate the hand pump continue un...

Страница 11: ...a new system altitude and press RUN F2 again 7 On completion of the test allow the altitude to decrease until it is below 1500 ft 8 Slowly open the manual let down valve to return the system to ground...

Страница 12: ...be potential free Figure 4 Airspeed Switch Test Connections 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pre...

Страница 13: ...ng the test 2 Ensure that positive pressure is selected and close the release valve 3 Close the manual let down valve 4 Zero the airspeed reading by pressing the ZERO key 5 Press the RUN key F1 and op...

Страница 14: ...layed operating data 6 Allow the airspeed to decrease until it is below 80 kts 7 Slowly open the manual let down valve to return the system to zero ground level SELECT TASK SELECT TASK ALTITUDE SWITCH...

Страница 15: ...1998 Baker Hughes Company English DPI610A Instruction Manual 13 8 To perform another leak test use the hand pump to generate a new airspeed and press RUN F2 again 9 To exit or cancel leak test mode p...

Страница 16: ...Copyright 1998 Baker Hughes Company 14 DPI610A Instruction Manual English...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...kerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 3...

Отзывы: