Copyright 2010 Baker Hughes Company.
English–TERPS Instruction Manual | 7
3.5.2 LabVIEW Driver
A National Instruments certified LabVIEW driver is available to download from the National
Instruments’ website:
http://sine.ni.com/apps/utf8/niid_web_display.model_page?p_model_id=30206
The driver includes a Virtual Instrument to discover a sensor’s address. Example code snippets
are supplied with the driver.
The LabVIEW driver is plug-and-play compatible with the USB sensor. A USB to RS
-
485 adapter,
or a USB to RS
-
232 adapter, can be used to communicate with the RS
-
485 and RS
-
232 sensors.
3.5.3 Mobile Application
A TERPS mobile application for Android™ is available for free download from Google Play™. The
TERPS application displays pressure from any USB TERPS sensor.
Barometric TERPS sensors can optionally display the following aeronautical parameters: QFE,
QNH, QFF and MSL.
3.6 Maintenance
3.6.1 Visual Inspection
Inspect the product for damage and corrosion. Any damage to the product must be assessed. If
the housing is no longer sealed against water and/or dust, the product must be replaced.
3.6.2 Cleaning
Clean the case with a damp lint-free cloth and mild detergent.
If the product has been in contact with hazardous or toxic materials, obey all the applicable
Control of Substances Hazardous to Health (COSHH) or Material Safety Data Sheet (MSDS)
references and precautions when handling.
3.6.3 Adjustment
For some models, the Output Calibration, Full-Scale and Offset settings can be adjusted. For
instructions see Chapter
6, “Programming Guide,” on page
Note:
Druck can provide a calibration service that is traceable to international standards.
3.7 Returned Goods Procedure
To repair or calibrate the sensor, return it to the applicable Druck Service Department.
Please contact our Service Department, and get a Return Authorization number.
Please supply these details:
•
Product (e.g. TERPS
8200 Pressure Sensor)
•
Pressure range
•
Serial number
•
Details of defect / work to be undertaken
•
Calibration traceability requirements
WARNING
High pressures and temperatures are dangerous and can cause
injury (Refer to pressure limits in the sales data sheet). Be careful when working
on components connected to lines that have high pressures and heat. Use the
applicable protection and obey all safety precautions.
WARNING
Output Calibration, Full-Scale and Offset adjustment may be
subject to state requirements for verification of metrological equipment.
Содержание Druck 8000 Series
Страница 1: ...Druck com 8000 8100 8200 8300 Series Trench Etched Resonant Pressure Sensor Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2010 Baker Hughes Company iv TERPS Instruction Manual English...
Страница 42: ...Copyright 2010 Baker Hughes Company 34 TERPS Instruction Manual English Appendix B RPS EEPROM Format...
Страница 43: ......