Table 4: Transducer signal settings
Transducer signal
parameters
Range
Default
value
Description
Acceleration limit
0 to 250 ft/s
(0 to 76 m/s)
15 ft/s
(5 m/s)
The E6: CYCLE SKIP error message appears when the calculated fluid
velocity changes by more than the programmed ACCELERATION LIMIT
value from one reading to the next. See Chapter 2 in the service manual
for a discussion of error codes.
Amplitude
discriminator low
0 to 100
14
The amplitude discriminator measures the transducer signal received
by the model XGF868i. The default value for the above parameter is
14, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error
message appears when the amplitude discriminator falls below the
programmed AMP. DISCRIM LOW value. See Chapter 2 in the service
manual for a discussion of error codes.
Amplitude
discriminator high
0 to 100
34
The amplitude discriminator measures the transducer signal received
by the model XGF868i. The default value for the above parameter is
34, and values from 0 to 100 are acceptable. The E5: AMPLITUDE error
message appears when the amplitude discriminator exceeds the
programmed AMP. DISCRIM HIGH value. See Chapter 2 in the service
manual for a discussion of error codes.
Delta T offset
-1000 to 1000
µsec
0 µsec
An offset between the upstream and downstream transit times is
specified at this prompt.
Skan T offset
–500 to 500
µsec
58 µsec
At this prompt, specify a time measurement offset that compensates
for any shift resulting from cross-correlation. Set to 0 for active skan T
offset.
% of peak
1 to 100%
50%
The percentage of peak used to calculate the transit times and Delta T
is specified at this prompt.
M>S switch
0 to 250 µsec
50µsec
If the burst mode is set to skan/measure (S/M), the meter switches from
skan to measure mode when Delta T is less than the M>S_switch value.
DO NOT change this value unless advised to do so by the factory.
# shifts
0 to 10
3
The number of shifts corresponds to the actual number of transmits per
cycle (number of signals added together in one direction to produce
an averaged signal for one interrogation of the fluid) and need only be
changed if the environment is very noisy or the acoustic signal is weak.
A divisor
0.1 to 10
2.5
A divisor is used to calculate the measure mode integrated threshold
level and is not normally changed.
# transmit pulses
1 to 16
4
# transmit pulses specifies the number of pulses in a burst. For difficult
conditions (i.e. long paths, high velocity or high temperature), settings
as high as 16 may be necessary.
T window (cycles)
0 to 1000
0
The XGF868i calculates the size of the transmit window based on pipe
size and fluid sound speed. However, for diagnostic purposes, the
window size may be reset.
R window (cycles)
10 to 128
10
The XGF868i calculates the size of the receive window based on pipe
size and fluid soundspeed. However, for diagnostic purposes, the
window size may be reset.
9
Содержание DigitalFlow XGF868i
Страница 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Страница 2: ...ii...
Страница 4: ...no content intended for this page iv...
Страница 10: ...x...
Страница 38: ...28...
Страница 46: ...no content intended for this page 36...
Страница 54: ...44...
Страница 60: ...50...
Страница 88: ...78...
Страница 94: ...84...