Bair Hugger Model 750 Temperature Management Unit Service Manual
16
Replacing the Air Filter
Service frequency
Every 6 months or 500 hours of use�
Tools/equipment
Replacement air filter
•
#2 Phillips screwdriver
•
warning
Do not attempt to clean the air filter as it may contain hazardous microorganisms� Discard the
filter in a manner consistent with institutional protocol for biohazardous material�
Method
Disconnect the temperature management unit from the grounded power source�
1�
Turn the temperature management unit upside down�
2�
Remove the 2 screws located in the louvered filter cover�
3�
Lift the filter cover off the temperature management unit�
4�
Remove the air filter and discard it�
5�
Place the new filter in the filter compartment with the gasket toward the filter ledge�
6�
Replace the filter cover with the louvers facing the back of the temperature management
7�
unit�
Replace the 2 screws in the filter cover�
8�