MID_PrimoAdvanced_Bedienungsanleitung_0704_e.doc
Parameter setting
Page 17/37
5.2.2.2 Pulse
output
The pulse value defines how many pulses per volume unit are dispensed. These
can be totalized and displayed as total flow volume.
A setting of 0,001 up to 10.000 pulse/volume unit is possible. A max. output
frequency of 10 kHz (10.000 pulses/sec) may not be exceeded. The program is
checking, which maximum pulse rate is possible at full scale and limits the input to
the maximum value.
Notice:
The volume unit depends on the flow unit (see units).
The pulse/break relation is approx. 1:1. At a setting of pulse width 50%, the pulse
width is automatically adjusted to each pulse frequency. The pulse width can be
programmed from 5 ms to 500 ms.
The program is checking, which maximum pulse length is possible at maximum
pulse output frequency (full scale) and allows no higher values.
The pulse can be inverted via the function pulse type.
-
+
Pulse output (active)
24 V DC
25 mA
max. 10 kHz
16
13
Pulse output (passive)
-
+
max. 30 V DC
250 mA
max. 10 kHz
16
15
1 second
50 ms
50 ms
Pulse width
50 ms
100 ms
100 ms
Pulse width
50%
1 second
Normal open
Normal
closed