MID_PrimoAdvanced_Bedienungsanleitung_0704_e.doc
Parameter setting
Page 16/37
5.2.1.3 Password
Notice:
Please contact the manufacturer at loss of the password.
The password consists of a number between 0 and 999. The value 0 is for „no
password protection active“. If a value bigger than zero is entered, the password
protection is activated. When calling the parameter setting mode again, the
password has to be entered.
5.2.2 Outputs
5.2.2.1 Analog
output
The following possible power ranges are available for the setting of the measuring
range 0 to 100% (= full scale):
Power output
0 to 20 mA
4 to 20 mA
0 to 10 mA
2 to 10 mA
Notice:
The power value is limited to 105% (21 mA resp. 10,5 mA) of the
full scale. If this value is exceeded, an overflow error (relay 3) is
indicated.
At bidirectional operation the flow direction is indicated via relay 1.
See also setting of the full scale.
-
+
Power output (active)
0/4...20 mA
0/2...10 mA
RL < 800 Ohm
1
3
Qv
Full scale
Flow direction
forward
10,5
mA
10 mA
Power output 10 mA
Flow direction
backward
2 mA
4...10 mA
0...10 mA
Full scale
Qr
0 mA
0 mA
Power output 20 mA
20 mA
21 mA
Flow direction
forward
4 mA
Flow direction
backward
4...20 mA
0...20 mA
Qr
Full scale
Full scale
Qv