3.2.3 Meter location
Caution:
•
Do not install the detector on the suction sides of pumps. This could
damage the liner (in particular PTFE liners).
•
Verify that the pipeline is always filled on the measuring point, if not - a
correct or accurate measurement is not possible.
•
Do not install the detector on the highest point of a pipeline system.
Gas accumulation may follow.
•
Do not install the detector in downcomer pipes with free outlet.
•
Do not install the detector on pipes with vibrations. If pipes are strongly
vibrating, make sure that detector and amplifier are separated
(separate version).