Product Overview 

PRODUCT OVERVIEW 

LCD Display

HR-LCD registers have a nine-digit Liquid Crystal Display (LCD) to show consumption, flow and alarm information .
There is no need to activate the display . The display automatically toggles between consumption (segmented leak detector in 

this mode), rate of flow and meter model .

OTE: 

N

Devices are shipped in storage mode so that a meter status alarm is not triggered . In storage mode, the meter model 

screen is displayed . 

Multiplier Value

Depending on the meter model, size and unit of measure, a multiplier value may also be shown . Multiply the displayed value 

by the multiplier value to calculate the reading to the nearest gallon, cubic foot, or cubic meter .  

Example: 123456 (displayed value) x 10 (multiplier value) = 1234560

Extended Visual

Reading Resolution

Typical Billing

Segments

Status

Indicators

IR Programming

Port

00/00

4-Digit

Date Code

E LCD

HR

Multiplier Value

(x10, x100, x1000, x10,000)

In 6-digit mode, the 

last digit is a dashed

line that becomes a

flow finder

Figure 1:  HR LCD register face

Visual Display

Units of Measure

The units of measure are factory-programmed and user-programmable . Options include U .S . gallons, Imperial gallons, cubic 

feet, cubic meters and liters .

9-Digit Totalization

The consumption display includes all nine digits and a decimal point (based on meter model, size and unit of measure) . The 

displayed value is the sum of the forward flow minus any reverse flow . This screen displays for 45 seconds .

Model 25 Disc Series Meter Calibrated in Gallons

4

GAL

Page 6 

April 2020

REG-UM-02466-EN-02

Содержание HR-LCD 4-20

Страница 1: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse REG UM 02466 EN 02 April 2020 User Manual...

Страница 2: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse Page ii April 2020 REG UM 02466 EN 02...

Страница 3: ...re 6 9 Digit Totalization 6 Rate of Flow 7 Meter Model Information 7 Installing the Register 7 Bayonet Mount 7 HR LCD 4 20 scaled unscaled 8 Wire Connections 8 Measurement Resolution 8 Specifications...

Страница 4: ...High Resolution LCD Registers HR LCD 4 20 scaled unscaled HR LCD Pulse Page iv April 2020 REG UM 02466 EN 02...

Страница 5: ...ity to operate the equipment DESCRIPTION High resolution registers are fully electronic solid state devices with no moving parts The devices come standard as factory programmed with the option for pro...

Страница 6: ...oot or cubic meter Example 123456 displayed value x 10 multiplier value 1234560 Extended Visual Reading Resolution Typical Billing Segments Status Indicators IR Programming Port 00 00 4 Digit Date Cod...

Страница 7: ...appears and Compound meters are indicated by a C See examples below Model 25 Disc Series Meter Calibrated in Gallons Model 450 Turbo Series Meter Calibrated in Gallons 2 in Low Side Compound Series M...

Страница 8: ...38 279 55 M70 1 1 10 100 46 773 349 884 12355 278 70 M120 1 1 2 0 10 1 10 23 866 178 533 6304 435 120 M170 2 0 10 1 10 14 565 108 950 3847 303 170 Recordall Turbo Series Size in Scaled pulse unit Unsc...

Страница 9: ...black wire negative Scaled digital output from the register has a default resolution of 1 10th of the register test circle resolution may vary in some cases The movement of the meter magnet is conver...

Страница 10: ...ies Size in Scaled pulse unit gal ft3 m3 LP 5 8 1 10 100 M25 5 8 1 10 100 M35 3 4 1 10 100 M40 1 1 10 100 M55 1 1 10 100 M70 1 1 10 100 M120 1 1 2 0 10 1 10 M170 2 0 10 1 10 Recordall Turbo Series Siz...

Страница 11: ...nt 100 mA Pulse Width 50 ms programmable 30 100 ms Power Input Voltage Range 9 50V DC supply Max Load Resistance Ohms 50 Ohms 50 Ohms supply voltage 9V Battery Lithium thionyl chloride AA cell fully e...

Страница 12: ...alarm remains active for 35 days The alarm automatically clears after 35 days if reverse flow condition has not recurred Suspected leak Register detects 24 hours without one 15 minute interval of no...

Отзывы: