Turbine replacement

TURBINE REPLACEMENT

The Model 1100 turbine flow meter uses wear-resistant moving parts to provide trouble-free operation and long service life . 

Designed for easy field service of a damaged flow meter, Model 1100 repair kits replace only the internal parts, not the entire 

flow meter . Repair parts use stainless steel alloys and tungsten carbide construction materials .
Each repair kit is factory calibrated to provide accuracy throughout the entire flow range . Each kit is complete and includes 

a new K-factor, which is the calibrated number of pulses generated by each gallon of liquid . Recalibration of the monitor or 

other electronics uses the K-factor to provide accurate output data .

OTE: 

N

If the meter repair kit part number ends in 

NCC

 (no calibration), it was not factory calibrated . For these repair kits, use 

the nominal K-factor .

Turbine Replacement Kit Part Number

Flow Meter Size

Replacement Kit Fits

Meter Part Number

Repair Kit Part Number

3/8 in .

B110-375, B110-375-1/2

B251-102

1/2 in .

B110-500, B110-500-1/2

B251-105

3/4 in .

B110-750, B110-750-1/2

B251-108

7/8 in .

B110-875

B251-109

1 in .

B111-110

B251-112

1-1/2 in .

B111-115

B251-116

2 in . Low

B111-121

B251-116

2 in .

B111-120

B251-120

3 in .

B111-130

B251-131

4 in .

B111-140

B251-141

6 in .

B111-160

B251-161

8 in .

B111-180

B251-181

10 in .

B111-200

B251-200

Standard Magnetic Pick-up

All Meter Sizes

B111109

Table 2: Repair kit part numbers

Turbine Assembly Removal

HIGH-PRESSURE LEAKS ARE DANGEROUS AND MAY CAUSE PERSONAL INJURY. MAKE SURE TO SHUT OFF FLUID FLOW 

AND RELEASE RESIDUAL PRESSURE IN THE LINE BEFORE ATTEMPTING TO REMOVE THE METER.

AVERTISSEMENT

LES FUITES À HAUTE PRESSION SONT DANGEREUSES ET PEUVENT CAUSER LE DOMMAGE CORPOREL. ASSUREZ-VOUS 

QUE LE FLUX DE FLUIDE A ÉTÉ COUPÉ ET DE LA PRESSION DANS LA LIGNE A ÉTÉ LIBÉRÉE AVANT D’ESSAYER D’ENLEVER 

LE MÈTRE.

Page 8 

November 2022

TUR-UM-00014-EN-09

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание Blancett 1100 Series

Страница 1: ...Turbine Flow Meter 1100 Series Turbine Meter TUR UM 00014 EN 09 November 2022 User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...7 Operational Startup 7 Turbine Replacement 8 Turbine Replacement Kit Part Number 8 Turbine Assembly Removal 8 Disassembly 9 New Turbine Kit Installation 10 Part Number Information 11 Troubleshooting...

Страница 3: ...y profile Fluid then passes through the turbine causing it to rotate at a speed proportional to the fluid velocity As each turbine blade passes through the magnetic field the blade generates an AC vol...

Страница 4: ...s INSTALLATION PRESSURE IN EXCESS OF ALLOWABLE RATING MAY CAUSE THE HOUSING TO BURST AND CAUSE SERIOUS PERSONAL INJURY AVERTISSEMENT LA PRESSION AU DESSUS DE L ESTIMATION PERMISE PEUT FAIRE CLATER ET...

Страница 5: ...AUSE DAMAGE ATTENTION DES DOMMAGES PEUVENT TRE PROVOQU S EN FRAPPANT UN M TRE VIDE AVEC UN JET D COULEMENT DE VITESSE LEV E Any restriction in the flow line may cause the liquid to flash If necessary...

Страница 6: ...eter 10 Pipe Diameters Figure 3 Meter installation with a bypass line 2 1 Electronic Flow Monitor Control Valve 5 Pipe Diameters Turbine Flow Meter 10 Pipe Diameters Figure 4 Meter installation withou...

Страница 7: ...T LIB R E AVANT D ESSAYER D INSTALLER LE M TRE DANS UN SYST ME ACTUEL After meter installation close the isolation valves and open the bypass valve Allow liquid to flow through the bypass valve for su...

Страница 8: ...Part Number Flow Meter Size Replacement Kit Fits Meter Part Number Repair Kit Part Number 3 8 in B110 375 B110 375 1 2 B251 102 1 2 in B110 500 B110 500 1 2 B251 105 3 4 in B110 750 B110 750 1 2 B251...

Страница 9: ...y Upstream Rotor Support Retaining Ring Retaining Ring Flow Meter Body Downstream Rotor Support Retaining Ring Groove Retaining Ring Groove Magnetic Pickup Figure 7 Component positions for B110 375 B1...

Страница 10: ...otors do not contain a cast or engraved arrow However a colored cap on the downstream side of the rotor shaft indicates flow direction Remove this cap before assembly noting flow direction 1 Install o...

Страница 11: ...4 515 6000 82 981 20 330 6 6 00 in 152 40 mm B311 066 1 1 2 in 38 1 mm 2 in grooved end 5000 15 180 56 8 681 4 515 6000 82 981 20 330 6 6 00 in 152 40 mm B111 120 2 in 50 8 mm 2 in female NPT 5000 40...

Страница 12: ...monitor Move to location where meter is full on startup or add down stream flow control valve Erratic indication at low flow good indication at high flow Rotor has foreign material wrapped around it C...

Отзывы: