Bacharach Refrigerant Recovery System 3600 Скачать руководство пользователя страница 16

Refrigerant Recovery System 

 

Model 3600 

 

16 

4.8.  DOT Recovery Cylinder Safety 

An  optional  80% recovery cylinder shutoff cord is available with the Model 3600.  If equipped, this cord 
connects to the recovery cylinder float switch and will automatically shut the Model  3600  off when the 
recovery cylinder becomes 80% full.   
 
When the Model 3600 is equipped with the 80% shutoff cord, Bacharach recommends that you use this 
cord for added safety.  If the Model 3600 is not equipped with an 80% shutoff cord, or, if you are using a 
recovery cylinder that does not have a float switch, then you must use a scale and the guidelines below to 
prevent overfilling of the cylinder. 

 

 

 

N

OTE

:

 

T

HE SHORTING CAP MUST BE USED IN ORDER TO OPERATE THE 

M

ODEL 

3600 

WITHOUT THE 

80%

 SHUTOFF 

CORD ATTACHED TO THE RECOVERY CYLINDER

 
Bacharach uses and recommends the Air Conditioning & Refrigeration Institute's (AHRI) guideline K for 
the safe filling and handling of used refrigerant.  This Publication is available from ARI. 
 

WARNING:  You must use a DOT recovery cylinder approved for use with R-410A refrigerant.   

This recovery system is equipped with a high-pressure shutoff switch set to turn the 
unit off at 550 psi.   Using any other cylinder may cause the pressure in the recovery 
cylinder to exceed the cylinder’s pressure rating. 

 
The following information provides the safe fill weights for used refrigerant based on the size of the 
container and is in accordance with Guideline K. 
 

Water Capacity 

Net Refrigerant 

Weight 

Gross Container Weight 

(Approx.) 

30 lbs 

24 lbs 

38 lbs 

50 lbs 

40 lbs 

59 lbs 

95 lbs 

76 lbs 

118 lbs 

145 lbs 

98 lbs 

153 lbs 

238 lbs 

190 lbs 

274 lbs 

 

 

 

 

 

 

 

Содержание Refrigerant Recovery System 3600

Страница 1: ...ing Vapors reduce oxygen available for breathing and are heavier than air Decomposition products are hazardous Liquid contact may cause frostbite All refrigerant containers equipment and hoses are und...

Страница 2: ...OF UNIT FEATURES 7 3 2 LED DISPLAY 8 CHAPTER 4 RECOVERY 9 4 1 HELPFUL HINTS FOR RECOVERING R 410A 9 4 2 VAPOR LIQUID RECOVERY PROCEDURES 9 4 3 VAPOR RECOVERY HOSE CONNECTION 11 4 4 LIQUID PUSH PULL RE...

Страница 3: ...tire operating instruction manual before use Failure to open the outlet port will cause permanent damage to the unit and void the warranty WARNING You must use a DOT recovery cylinder approved for use...

Страница 4: ...ndensing ensures high recovery rates even on the hottest days Self purging valve completely purges the condenser and tubing eliminating cross contamination when switching between refrigerants Solid st...

Страница 5: ...t 18 inches above the floor This equipment should not be used in the vicinity of spilled or open containers of flammable materials 5 DISCONNECT UNIT BEFORE SERVICING Electrical shock hazard may be pre...

Страница 6: ...in enclosed and low lying areas Avoid contact of liquid with eyes and prolonged skin exposure Wear safety goggles and protective gloves Do not apply open flame or heat unit above 125 F Do not allow re...

Страница 7: ...port It points DOWN when it is open and to the left when CLOSED 6 Inlet Port Refrigerant hose connection for incoming refrigerant vapors 7 Outlet Port Refrigerant hose connection for outgoing refrige...

Страница 8: ...power switch is ON and the compressor is running Red High Pressure Condition Indicates that there is a high pressure condition that must be corrected The internal high pressure switch activates at 55...

Страница 9: ...ns Connect refrigerant hoses to recovery cylinder the unit and A C system as shown in Figure 1 NOTE You must use an external filter drier in the suction line in order to meet the conditions of the uni...

Страница 10: ...NOTE A small amount of inlet pressure may be needed to reset the vacuum switch this is normal Red a Indicates that there is a high pressure condition that must be corrected Check for restrictions or a...

Страница 11: ...very Hose Connections Use only approved DOT recovery cylinders Follow precautions and warnings listed on the cylinder If servicing a system with a compressor burn out use a high acid type filter drier...

Страница 12: ...ng non compressible fluids such as liquid refrigerant however the 3600 has been designed with a special head and valves that enable it to handle liquid directly This feature makes the 3600 easier to u...

Страница 13: ...m switch to the OFF position NOTE Liquid push pull operations do not pull a vacuum on a system You must perform a vapor recovery operation in order to pull the system down to the required vacuum level...

Страница 14: ...Connections Use only approved DOT recovery cylinders Follow precautions and warnings listed on the cylinder If servicing a system with a compressor burn out use a high acid type filter drier and do n...

Страница 15: ...mpressor properly equalizes 2 Turn the system switch to the ON position The LED status will be green NOTE If the compressor fails to start cycle the purge valve from the RECOVERY to PURGE position to...

Страница 16: ...RATE THE MODEL 3600 WITHOUT THE 80 SHUTOFF CORD ATTACHED TO THE RECOVERY CYLINDER Bacharach uses and recommends the Air Conditioning Refrigeration Institute s AHRI guideline K for the safe filling and...

Страница 17: ...20 volt power source Fan Will Not Start Green LED Defective fan Replace fan Obstruction in fan blades Remove obstruction NONE System switch is off Turn system and fan switch to on position Unit Starts...

Страница 18: ...Refrigerant Recovery System Model 3600 18 CHAPTER 6 ELECTRICAL SCHEMATIC...

Страница 19: ...00 2024 0050 VALVE CHECK 1 4 1 00 2024 0270 VALVE 3 WAY BALL 1 8 FEM NPTF DROP 2 00 2030 0031 CONDENSER ASSY MODEL 3600 1 00 2035 0380 SUCTION HOSE 1 4 FLR X 1 4 FLR BLUE 1 00 2035 0390 DISCHARGE HOSE...

Страница 20: ...del 3600 20 World Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 1 800 736 4666 Web site www MyBacharach com E mail help MyBacharach com PRINTED...