![Bacharach HGM300 Скачать руководство пользователя страница 6](http://html1.mh-extra.com/html/bacharach/hgm300/hgm300_power-supply-replacement_3303744006.webp)
HGM300 Power Supply Replacement
6
P/N 3015-5686 Rev 0
Figure 4. Comparison of Old Style and New Style Power Wiring and Grounding
2.4. REQUIRED TOOLS
•
New Power Supply printed circuit board (P/N 3015-5675)
•
HGM300 Operations Manual (P/N 3015-4148)
•
Phillips head screwdriver
SECTION 3: PROCEDURE
Step
Description
1
Refer to the HGM300 Operations Manual (P/N 3015-4148) originally supplied with your HGM300. It
contains warnings, wiring, and operation information that is supplemental to this replacement
procedure.
2
Remove power that supplies the HGM300. Perform appropriate lock-out/tag-out procedures to ensure
power remains off during the power supply board replacement process.
3
Open the HGM300 enclosure and locate the Power Supply board to be replaced. See Figure 2.
4
Remove all connectors from the Power Supply board, noting their orientation. This includes the two
input power connectors, the power supply connector to the pump, and the two power supply
connectors to the main board.