Operation
Halogen Monitor
10
Instruction 3015-4584 Rev 9
During normal operation, the
Data Display
screen shows when the monitor is performing the following
two functions:
MEASURE
indicates that the monitor is actively measuring gas. In the screen shown above, 10 ppm of
R-134A gas is currently being detected, and that a peak measurement of 20 ppm has occurred sometime
in the past. The measurement cycle runs for 4 minutes.
To reset the peak value to zero, press the
ENTER
button.
2.9.4 Function Screens
The Function
screens are used to set up the monitor and display stored data.
From the
Data Display
screen, press both
the
Keypad
Left
and
Right
buttons at the same time to
display the first
Function Menu
screen. Next, use the
Keypad
buttons to move the arrow (>) until it
points to the desired function, and then press the
ENTER
button to select that function.
Once a function has been selected, use the
Keypad
to scroll through the displayed data or to change a
parameter associated with that function. Press
ENTER
to save any newly entered parameters. Press the
ESC
button to return to the previous screen without saving.
Note that if no buttons are pressed within 90 seconds after selecting a function, the unit returns to the
Data Display
screen.
FILLBAG
– Press
ENTER
to manually fill the purge-air bag.
WARNING:
Fill the purge-air bag in a clean, fresh-air
environment that is at the same temperature and humidity as
the area being checked for gas. DO NOT fill the bag in an
area that is contaminated with gas.
EMTYBAG
– Press
ENTER
to manually empty the purge-air bag. Use
this function to drain the bag if it becomes contaminated with gas.
ALRMLVL
– Detected gas level that causes the front panel
ALARM
(red) light to turn ON and the audible alarm to sound (if AUDALRM
function is set to ALARM or ALARM/FAULT).
AUDALRM
– The internal audible alarm is programmable to function as
follows:
•
OFF
– Audible alarm is disabled.
•
ALARM
– Alarm sounds when detected gas level exceeds the
ALRMLVL level as evidenced by the front panel ALARM (red) light
turning ON.
•
ALARM/FAULT
– Alarm sounds when either the detected gas level exceeds the ALRMLVL value, or if
a system fault occurs, as evidenced by the front panel ALARM (red) light or FAULT (yellow) light
turning ON.
•
LEAK DETECTOR
– When selected, the monitor will produce a leak detector type of audible
feedback. At zero ppm there will be a tic every 2 seconds with the tic rate increasing up to 20 tics per
second when the ppm level reaches the ALRMLVL setting.
>FILLBAG EMTYBAG
ALRMLVL AUDALRM
>PPM LOG LOG INT
FAULTS DIAG
>SILENCE P-CHK
GAS CLOCK
FILLING BAG...
DRAINING BAG...
AUDIBLE ALARM IS
ALARM
EDIT ALARM LEVEL
300 ppm
>2ndUNIT PRINT
PLOT PC-DUMP