background image

H25-IR PRO External Filter Retrofit 

 

 

P/N: 3015-9004 Rev 0 

5

 

Step

 

Description

 

Disconnect the AC power 

entry module wires, noting 

the wire color and position.  

 

Remove the two lower chassis 

screws.  

Carefully pull the rear plate 

away from the back of the 

chassis.  

10 

Locate the exhaust tube port 

inside the instrument, where 

it exits. Pull the tubing off the 

barbed fitting. The rear plate 

will now be free from the 

instrument.  

2

.

2

Part Transfer to New Plate

 

Step

 

Description

 

Using the tools described earlier, remove and reinstall the speaker, RS-232, exhaust port barbed fitting, and 

audio PCB onto the new rear plate. Connect the speaker as required. 

The AC power entry module is removed by pressing the springs on each side inward until they are flat. Do this 

while pressing down on the module to aid release. The terminal blocks removed earlier may be used as a 

support for the plate while the power entry module is removed. As soon as one side is released the other side 

will release easily. 

Insert the power entry module in the new rear plate, noting that the power switch is to the right, and the 

ground pin is up. 

2

.3. 

Installation of New Plate

 

Step

 

Description

 

Installation is generally the reverse of the removal steps. 

It is recommended that the exhaust tube be connected first. 

Align the new rear plate to the instrument PCB terminal strip. Insert the terminal blocks to aid holding the plate 

in place.  
NOTE: Ensure the new filter tubing is not pinched, kinked, or trapped between any components. 

Reconnect the audio PCB, RS232, and power entry module wires. 

2

.4. 

Tubing Changes

 

Step

 

Description

 

Locate the 25 mm diameter filter and sample inlet tubing. The tubing is 1/8” (3 mm) silicone. 

Disconnect the filter from the larger tubing nearest the main PCB. 

Содержание H25-IR PRO

Страница 1: ...External Filter Retrofit Instruction 3015 9004 Revision 0 September 2014 Product Leadership Training Service Reliability...

Страница 2: ...or to include the newer external filter Figure 1 1 Back Panel Comparison of Previous and Current Generations of the H25 IR Pro 1 2 Required Tools The following items are required to complete the retro...

Страница 3: ...nel includes 1 new filter P N 3015 5816 1 3015 5818 5 pack of additional new filters pack of 5 of 3015 5816 17 in 0003 6372 Silicon tubing 1 16 ID x 1 8 OD for inlet line 7 in Silicon tubing 1 16 ID x...

Страница 4: ...Unplug the probe assembly 3 Remove the front and rear bezel by gently pulling at the left or right center point 4 Remove the top chassis screws at the front and back Pry up on the chassis lid and remo...

Страница 5: ...n the module to aid release The terminal blocks removed earlier may be used as a support for the plate while the power entry module is removed As soon as one side is released the other side will relea...

Страница 6: ...H25 IR PRO External Filter Retrofit 6 P N 3015 9004 Rev 0 Figure 2 1 Internal Tubing Comparison of Previous and Current Generations of the H25 IR Pro...

Страница 7: ...cal connections are not damaged 5 Twist the tubing on the tee next to where the internal filter was connected so that the tubing stub is pointing upward See the photos below 6 Locate the 7 178 mm long...

Страница 8: ...disconnect the top tubing from the fitting Install the new filter noting the proper orientation shown below 4 Repeat for the lower tubing Dispose of the used filter 5 Install the filter in the opposi...

Страница 9: ...ting one of the following Bacharach Service Centers United States Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5051 Fax 724 334 5723 Email help mybacharach com Canada Bac...

Страница 10: ...rnal Filter Retrofit 10 P N 3015 9004 Rev 0 Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Website www mybacharach com E mail help mybacharach com Printed in U S A Registered Trademark of...

Отзывы: