Instruction 0011-9026
Page 10
To make a test with the FYRITE, the metal Sampling Tube at one end of Sampling
Assembly Hose is inserted into the gas to be analyzed. The rubber Connector Plug at
the other end of the Rubber Hose is then
fi
rmly pressed down on the spring-loaded
Valve of the instrument (See Figure 12). This simultaneously opens a passage into the
Top Reservoir and seals off the center bore.
Next, a sample of the gas is pumped into the Top Reservoir by squeezing the rub-
ber Aspirator Bulb. At least 18 bulb squeezes are required to assure that the rubber
Sampling Hose and the Top Reservoir of the FYRITE are thoroughly purged of the
previously analyzed sample. During the 18th Aspirator Bulb squeeze and immediately
before releasing Bulb, the
fi
nger is lifted from the Connector Plug, which automati-
cally returns the Plunger Valve to upper position against its top seat.
Refer to Figure 13. With the Valve in this position, the
fi
nal 60 cc of the gas sample is
locked into the FYRITE and the Top Reservoir is opened to the center bore so that the
gas sample can pass to the absorbing
fl
uid. The FYRITE is then inverted, forcing the
gas sample to bubble through the absorbing solution which absorbs either CO
2
or O
2
depending on the type FYRITE in use.
Refer to Figure 14. The instrument is then turned and held upright. Absorption of the
CO
2
or O
2
by the absorbing
fl
uid creates a suction, which causes the Diaphragm at the
bottom to
fl
ex upward permitting the level of the absorbing
fl
uid to rise in the center
tube an amount proportional to the gas absorbed. The scale indication correspond-
ing to the top of the
fl
uid column is then read as a percentage of gas absorbed by the
FYRITE
fl
uid.
5.0 APPLICATION INFORMATION
FYRITE Analyzers are available for either Carbon Dioxide or Oxygen analysis, and
each of these models is made in the three scale ranges shown in Section 5.3. All six
instruments are similar in appearance and size, but they differ in important construc-
tion details, as well as in absorbing
fl
uids.
5.1 FYRITE Absorbing Fluid
FYRITE absorbing
fl
uid is selective in the chemical absorption of carbon dioxide or ox-
ygen respectively. Therefore, the FYRITE’s accuracy is well within the range required
for industrial and professional applications, and does not depend upon complicated
sequential test procedures. Further more, the FYRITE readings are unaffected by the
presence of most back ground gases in the sample.
The appearance of a dark red
fl
uid
fl
oating on top of CO
2
FYRITE solu tion is entirely
normal. The darker
fl
uid consists of a small amount of material added to prevent
excessive foaming at the meniscus and improves readability.
5.1.1 FYRITE Refi ll Fluid (Storage Life, etc.)
FYRITE re
fi
ll bottles should be stored at room temperature (70° F [21° C]) in the car-
ton provided, and placed in use within one year.
NOTE:
Before opening re
fi
ll bottle, read the label and make
certain to observe all precautions.