To make a test with the FYRITE, the metal Sampling Tube at one end of Sampling 

Assembly Hose is inserted into the gas to be analyzed. The rubber Connector Plug at 

the other end of the Rubber Hose is then 

fi

 rmly pressed down on the spring-loaded 

Valve of the instrument (See Figure 12). This simultaneously opens a passage into the 

Top Reservoir and seals off the center bore.

Next, a sample of the gas is pumped into the Top Reservoir by squeezing the rub-

ber Aspirator Bulb. At least 18 bulb squeezes are required to assure that the rubber 

Sampling Hose and the Top Reservoir of the FYRITE are thoroughly purged of the 

previously analyzed sample. During the 18th Aspirator Bulb squeeze and immediately 

before releasing Bulb, the 

fi

 nger is lifted from the Connector Plug, which automati-

cally returns the Plunger Valve to upper position against its top seat.

Refer to Figure 13. With the Valve in this position, the 

fi

 nal 60 cc of the gas sample is 

locked into the FYRITE and the Top Reservoir is opened to the center bore so that the 

gas sample can pass to the absorbing 

fl

 uid. The FYRITE is then inverted, forcing the 

gas sample to bubble through the absorbing solution which absorbs either CO

2

 or O

2

 

depending on the type FYRITE in use.

Refer to Figure 14. The instrument is then turned and held upright. Absorption of the 

CO

2

 or O

2

 by the absorbing 

fl

 uid creates a suction, which causes the Diaphragm at the 

bottom to 

fl

 ex upward permitting the level of the absorbing 

fl

 uid to rise in the center 

tube an amount proportional to the gas absorbed. The scale indication correspond-

ing to the top of the 

fl

 uid column is then read as a percentage of gas absorbed by the 

FYRITE 

fl

 uid.

5.0   APPLICATION INFORMATION

FYRITE Analyzers are available for either Carbon Dioxide or Oxygen analysis, and 

each of these models is made in the three scale ranges shown in Section 5.3. All six 

instruments are similar in appearance and size, but they differ in important construc-

tion details, as well as in absorbing 

fl

 uids.

5.1   FYRITE Absorbing Fluid

FYRITE absorbing 

fl

 uid is selective in the chemical absorption of carbon dioxide or ox-

ygen respectively.  Therefore, the FYRITE’s accuracy is well within the range required 

for industrial and professional applications, and does not depend upon complicated 

sequential test procedures. Further more, the FYRITE readings are unaffected by the 

presence of most back ground gases in the sample.

The appearance of a dark red 

fl

 uid 

fl

 oating on top of CO

2

 FYRITE solu tion is entirely 

normal. The darker 

fl

 uid consists of a small amount of material added to prevent 

excessive foaming at the meniscus and improves readability.

5.1.1   FYRITE Refi ll Fluid (Storage Life, etc.)

FYRITE re

fi

 ll bottles should be stored at room temperature (70° F [21° C]) in the car-

ton provided, and placed in use within one year.

NOTE:

 

Before opening re

fi

 ll bottle, read the label and make 

certain to observe all precautions.

Содержание FYRITE CO2

Страница 1: ...Instruction 0011 9026 Page 1 Printed in U S A Registered Trademarks INSTRUCTION 0011 9026 FYRITE Gas Analyzer CO2 and O2 Indicators Operation Maintenance Rev 12 December 2012 20 16 12 8 4 CO2 0...

Страница 2: ...graph 1 does not apply to parts the Operating Instructions designate as having a limited shelf life or as being expended in normal use THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED...

Страница 3: ...should always be in the bottom reservoir before and during sample introduction to the FYRITE When venting the FYRITE to atmosphere hold the instrument in the vertical position and at a location away...

Страница 4: ...n Efficiency 13 5 5 O2 FYRITE Only Combustion Testing 13 5 5 1 Percent O2 Shows Volume of Excess Combustion Air 13 5 5 2 Too Much Excess Air Means Inefficient Combustion 13 5 5 3 What is Proper O2 Val...

Страница 5: ...molded plastic cap that contains a double seated plunger valve A spring holds this valve against a carefully finished seat in the top cap providing a perfect seal which makes the instrument spill proo...

Страница 6: ...Assembly Always check sampling assembly for proper leak free operation as outlined in Section 6 7 before proceeding with test d Make certain wool in Filter Saturator Tube is moistened with water if u...

Страница 7: ...r inhalation overexposure If irritation develops move victim to fresh air For swallowing Do not induce vomiting Give two glasses of water or milk if conscious and not convulsing Get immediate medical...

Страница 8: ...gure 4 1 Hold upright Fig 1 and away from face Depress Plunger Valve momen tarily to vent FYRITE and release 2 Invert FYRITE Fig 2 to absorb O2 drawn into FY RITE whenever Plunger Valve is depressed H...

Страница 9: ...nector tip Pump sample by squeez ing and releasing as pirator bulb 18 times During 18th bulb squeeze with bulb held deflated release connector tip and Plunger Valve 5 Holding FYRITE upright Fig 5 loos...

Страница 10: ...point corresponding to top of the fluid column CO2 FYRITE OPERATION O2 FYRITE OPERATION RED FLUID BLUE FLUID 7 Absorb sample gas into FYRITE by inverting until fluid drains into top reservoir Fig 7 Th...

Страница 11: ...nts In the event of a spill read In structions in Section 2 2 FYRITE fluid handling precautions 3 1 Determining CO2 and O2 FYRITE Fluid Strength FYRITE Fluid strength can be conveniently checked immed...

Страница 12: ...e temperature variations are involved Temperature stabilization can also be accelerated by invert ing FYRITE frequently 3 3 1 Temperature of Gas Sample The gas sample as it enters FYRITE should be coo...

Страница 13: ...er valve is released the gas sample is locked into FY RITE and the top reservoir is opened to center bore so that gas sample can pass through absorbing fluid Figure 14 Absorption of gas sample by flui...

Страница 14: ...absorbing fluid creates a suction which causes the Diaphragm at the bottom to flex upward permitting the level of the absorbing fluid to rise in the center tube an amount proportional to the gas absor...

Страница 15: ...is now upright and the O2 fluid is filling the instrument see Figure 21a Immediately install plastic Top Cap Ring without delay This method helps to minimize oxidation and improve the life of the flu...

Страница 16: ...Checking CO2 in sewage plant digesters Oxygen tests in connection with oxygen and gas anesthesiology 5 4 CO2 FYRITE Only Combustion Testing 5 4 1 Percent CO2 Shows Volume of Excess Air Refer to Figure...

Страница 17: ...ucts Testing for smoke CO or combustibles is important since in extreme fuel rich firing high CO2 values also occur see Figure 16 5 4 4 How To Calculate Combustion Efficiency Measure percent CO2 in fl...

Страница 18: ...th excess air Gaseous fuels should also be checked for presence of carbon monoxide CO in the flue gases and adjusted for either no CO or to conform to applicable safety regulations 5 5 5 Where is Perc...

Страница 19: ...on between Oxygen CO2 and excess air in flue gases for Natural Gas and Fuel Oil THEORETICAL EXCESS AIR CURVE PERCENT EXCESS AIR STOICHIOMETRIC MIX Figure 16 CO2 measurements alone do not determine com...

Страница 20: ...when the sample no longer contains the interfering substance They can therefore be compensated for by saturating the FYRITE solution with samples containing the mechanical interferant as long as the...

Страница 21: ...ion NOTE The FYRITE is calibrated to analyze gases which are normally saturated with moisture Failure to moisten wool packing in filter saturator tube with water will cause FYRITE to indicate slightly...

Страница 22: ...1829 0 2 0 3 0 2 7000 2134 0 3 0 3 0 3 8000 2438 0 4 0 4 0 3 9000 2743 0 4 0 5 0 4 10 000 3048 0 5 0 6 0 5 11 000 3353 0 6 0 7 0 5 12 000 3658 0 7 0 8 0 6 13 000 3962 0 7 0 8 0 6 14 000 4267 0 8 0 9...

Страница 23: ...om Plunger Valve while still exhaling breath Perform steps outlined below to absorb exhaled breath into FYRITE fluid 1 Absorb sample gas into FYRITE by inverting until fluid drains into top reser voir...

Страница 24: ...NGER VALVE ASSEMBLY OVAL HEAD SCREWS TOP CAP RING WITH GASKET 20 16 12 8 4 O2 0 Figure 18 Increasing FYRITE fluid to proper level Figure 19 Decreasing FYRITE fluid to proper level Figure 20 Removing t...

Страница 25: ...When it is desirable to check fluid strength before taking the FYRITE to a location where it will be used it can be conveniently tested on a sample of atmospheric air which contains 20 9 O2 NOTE Make...

Страница 26: ...into FYRITE Reinstall plastic Top Cap Assembly and metal Top Cap Ring CO2 O2 FYRITEs Make certain that Top Cap Ring Gasket 11 0109 Figure 33 is in place between metal Top Cap Ring and plastic Top Cap...

Страница 27: ...p Gasket used in 7 20 21 Models only 20 16 12 8 4 CO 2 0 Figure 21a Alternate filling meth od Invert FYRITE with bottle in place to upright position Figure 22 Examining top gasket for warpage Figure 2...

Страница 28: ...ss Plunger Valve against its spring limit and strip old Gasket from the end of Plunger Valve Before assembling new Gasket wet inside surface of Tip Gasket then force it over the end of Plunger Valve a...

Страница 29: ...26 Page 25 Figure 25a Replacement Diaphragm properly installed Figure 25 Locating Bezel Rubber Gaskets Figure 26 FYRITE Sampling Assembly Locating Inlet Outlet Check Valves 20 16 12 8 4 CO2 0 LETTERIN...

Страница 30: ...automobile trunk over the Summer is not recommended When stored unused over the Summer it is good practice to drain old fluid and replace with fresh charge fluid when needed in the Fall FYRITE Refill...

Страница 31: ...BER CONNECTOR TIP HOLD FINGER HERE CHECK FOR BULB INFLATION AFTER SQUEEZING SAMPLING TUBE END HOLD FINGER HERE Figure 27 Testing Sampling Assembly outlet side for leaks Figure 28 Testing Sampling Asse...

Страница 32: ...Instruction 0011 9026 Page 28 Figure 30 Wrapping Replacement Filter Material Figure 31 Wet Filter Material then squeeze out excess water Figure 32 Installing wetted Filter Material into Saturator Tube...

Страница 33: ...21 Models only 1 11 0140 FYRITE Body All Models except 7 0 CO2 O2 1 11 0154 FYRITE Body 7 Models only 1 11 0062 Filler 60 FYRITE Adapter Plug 1 05 5134 Bezel Gasket O Ring 2 05 5169 O Ring Filler Oute...

Страница 34: ...parts breakout 11 0019 11 0026 11 0188 11 0188 11 0102 11 0136 11 0109 11 0132 11 0062 05 5169 05 5155 11 0102 11 0143 7 20 and 21 Models Only 11 0110 11 0110 11 0020 11 0140 11 0154 05 5134 1 1 0 1...

Страница 35: ...011 0122 0011 0180 0011 0130 0011 0119 0011 0138 0011 0138 0011 0120 0011 0152 7 2 PARTS LIST FOR FYRITE SAMPLING ASSEMBLIES STANDARD SAMPLING ASSEMBLY Part No 11 7029 Gases saturated with water vapor...

Страница 36: ...der with Orifice 1 11 0120 10 0022 Aspirator Bulb 1 11 0138 10 0023 Inlet Outlet Valve Red 2 11 0119 10 0024 Rubber Tubing 3 length 1 11 0180 10 0025 Filter Nipple with Bushing 2 11 0130 10 0026 Filte...

Страница 37: ...Instruction 0011 9026 Page 33 Notes...

Страница 38: ...Instruction 0011 9026 Page 34 Notes...

Страница 39: ...Instruction 0011 9026 Page 35 Notes...

Страница 40: ...nstruction 0011 9026 Page 36 Bacharach Inc 621 Hunt Valley Circle New Kensington PA 15068 Phone 724 334 5000 Fax 724 334 5001 Toll Free 800 736 4666 Web www mybacharach com E mail help mybacharach com...

Отзывы: