MODEL 300 COMBUSTION ANALYZER
Instruction 24-9223
3-11
OK
W*CO
E*O2
E*AM
-----
Remove and then reapply power to
the analyzer. Allow pump to run
5 minutes (longer if necessary) while
sampling fresh air to purge the
instrument of CO. Then press CAL.
Adjust potentiometer R9. Refer to
Section 4.3.
Replace CO sensor. Refer to Section
5.2.
Remove and then reapply power to
the analyzer. Allow pump to run 5
minutes (longer if necessary) while
sampling fresh air to purge instru-
ment of O
2
. Then press CAL.
Replace O
2
Cell. Refer to Section 5.1.
Allow analyzer to stabilize at a
temperature between 32° and 122° F
(0° and 5° C).
Adjust potentiometer R24. Refer to
Section 4.2.
TABLE 3-1. Basic Error, Warning and Overrange Displays
Display
Shows
Display Meaning
Remedy
No errors, warnings or
overranges to report.
Warning, CO level over
115 ppm after calibra-
tion.
Warning, CO span
potentiometer R9
misadjusted.
Warning, CO sensor
bad.
Error, O
2
Cell output
voltage is outside its
normal operating range.
Error, O
2
Cell bad.
Error, ambient tem-
perature below 32° or
above 122°F (0° &
50°C).
Error, ambient tem-
perature calibrate
potentiometer R24
misadjusted.