5.0 CALIBRATION & MAINTENANCE
Important:
Fresh batteries should be installed, and the unit
allowed to sta bi lize at room temperature for at least two
hours before proceeding with calibration. To main tain ac-
cu ra cy as listed in the Technical Characteristics Sec tion of
this manual, the standards used must be at least four times
as accurate as stated accuracy of the Fyrite Tech.
5.1 Entering the Calibration Mode and
Testing the Display Segments
1. With the analyzer turned OFF, place the unit in fresh, ambient air;
then press and hold down the
ENTER
button.
2. Press the
I/O
button and release it. Observe that all LCD seg ments
are turned ON.
3. Release the
ENTER
button. Observe the unit’s model number and
soft ware version are displayed. The word “CAL” is then displayed
while the unit counts down from 60 seconds.
.
At the end of 60 seconds, the fi rst calibration screen is automati-
cally displayed.
Note:
During calibration, the
▲
and
▼
buttons are used to
in crease or de crease a displayed calibration value.
ENTER
is used to store the new value and move to the next screen.
Hold
ENTER
down for 2 seconds to exit the Calibration
Mode and directly enter the Run Mode.