Fyrite INSIGHT
36
4. Pressing the
RUN/hOlD
button freezes all readings, stops the pump and
displays the Combustion-Test HOLD screen. Scroll to view all test values
at the moment the
RUN/hOlD
button was pressed. Pressing
RUN/hOlD
again restarts the pump and resumes testing.
Display Name
Description of Measurement or Calculation
O
2
% Oxygen
CO
ppm Carbon Monoxide
LAMBDA Excess air
CO
2
% Carbon Dioxide
CO
2
Max
% CO
2
in flue gas assuming perfect combustion
T-STK
Stack Temperature
T-AIR
Primary/Ambient Air Temperature
qA
Stack loss
Eta
Efficiency referenced to the lower heating value
Eff
% Efficiency referenced to the higher heating value
CO/CO2
Carbon Monoxide/Carbon Dioxide ratio
CO(n)
Carbon Monoxide content referenced to % of O2 concen
-
tration
AVG SMOKE
Average of 3 manually entered smoke numbers
OIL DERIVE
Manually entered Yes or No
BOILER TEMP
Boiler temperature manually entered
4.6 Making a Draft / Pressure Measurement
The difference in pressure (ΔP) between two areas can be measured by
using the analyzer’s two pressure ports and the Pressure screen. By using
the -ΔP port as the reference, the pressure applied to the +ΔP port will be
displayed on the Pressure screen as the differential pressure between the
two ports. Perform a draft / pressure measurement as follows:
1. Power on the analyzer and allow it to complete its warm-up cycle.
2. Press the
MENU (F2)
button to access the Main Menu.
3. Use the ▲▼ buttons to select Pressure. Press
ENTER
to display the
Pressure screen.