0019-9376
Revision 0
54
Monoxor
®
XR User Manual
5.4.2 B-Smart
®
CO Sensor Replacement
1. Enter the CALIBRATION MENU. Note that this requires password validation
(see section
3.8)
.
2. Use the UP ( ) and DOWN ( ) arrow buttons to select B-Smart
®
. Press ENTER to display
the B-Smart
®
code screen.
3. Use the UP ( ) and DOWN ( ) arrow buttons to enter the 10-digit alphanumeric code
supplied with the pre-calibrated B-Smart
®
sensor. Use the LEFT ( ) and RIGHT ( ) arrow
buttons to move the cursor across the screen. Press ENTER.
Figure 5-4:
CO Sensor Replacement Screens
NOTE:
If the correct code was entered, the analyzer accepts it and returns
to the CALIBRATION MENU. If an incorrect code was entered, the screen will
display “Invalid Code.” Check to make sure the correct code has been entered
If the problem persists, contact your nearest Bacharach Service Provider.
NOTE:
B-Smart
®
codes can be entered through the Fyrite
®
User Software
(FUS)
NOTE:
Installing a B-Smart
®
sensor forces the instrument to perform a zero
function
(either manual or automatic)
NOTE:
Bacharach offers a convenient Exchange Program
(where available)
that
allows the customer to regularly receive pre-calibrated replacement sensors that
include a code that can be entered into the analyzer for a quick convenient setup.
Contact Bacharach customer service for more details about this program.