Always disinfect the nebulizer set, mouthpiece and other nebulizer parts before using the nebulizer for the first time, after the last treatment of the
day, or daily during normal use.
Disconnect the Air Tubing and set aside.
Clean the casing of the nebulizer and air connectors by using a soft cloth moistened with alcohol. Do not use abrasive cleaners. Dry the casing
immediately using a soft clean cloth.
Disassemble mouthpiece parts or mask from the nebulizer accessories set
Using a clean container or bowl, soak all accessories item except Air Tubing in 75% alcohol or 3% hydrogen peroxide for 5 minutes or use a
medical bactericidal disinfectant available through your provider.
With clean hands, remove the accessories item from the disinfectant solution, rinse under clean water.
Air dry or hand dry in a clean environment.
When dry, reassemble the nebulizer and store for the next use.
The nebulizer has an Air Inlet filter to remove airborne contaminants. On average, the Air filter requires regular replacement after every 60days. The
nebulizer comes with a package of extra filters for this purpose. Check the Air Inlet filter at least once every two weeks. Following the below
mentioned steps to check and replace the Air Filter when needed.
1. Hold the Air Inlet filter cover ”A3” with your fingers and turn anti -clockwise to unlock it and then pull out.
2. Now you can see the Air Filter in the filter compartment. Remove the Air Filter from the filter compartment (if needed, you may use a toothpick to
remove the old filter).
3. If change is required, insert a new Air Filter into the filter compartment and discard the dirty Air Filter, otherwise reinsert the original Air Filter into
the filter compartment.
4. Insert the filter cover in position and turn clockwise to lock it in position.
1. If the cord is damaged, it must be replaced by an Authorized Service Centre.
2. If the nebulizer fails to operate, unplug the unit for 15minutes, plug in and try again. If the problem persists, check the “TROUBLE SHOOTING”
section. If problem still persists, then contact Authorized Service Centre.
Nebulizer not
work at all
Check that the power cord is connected to the nebulizer and
plugged in the mains. Check the fuse in the plug if there is any.
Nebulizer is
noisy
Check that the filter is locked correctly
Dimensions
Approx: 220(L) X 160(W) X 116(H) mm
Weight
1.8 kg
Electrical
220-240V ~ 50Hz
Power
80VA, MAX 2.5A
Continuation of
Operation Time
≦
30 mins / operation
Nebulizer Rate
0.3 - 0.4ml / min.
Nebulizer rate may vary with different medication used. Especially
suspensions and high viscosity medications may have a big
influence on this specification. See drug supplier’s data sheet for details
Particle Size
(NaCl Saline)
0.0 - 5.0
μ
m
>
86%
Particle size may vary with different medication used. Especially
suspensions and high viscosity medications may have a big
influence on this specification. See drug supplier’s data sheet for details
Medication
Cup Capacity
6ml ( cc) max.
Appropriate
Medication Qty
2ml - 6ml (cc)
Suction
Pressure
11 PSI max.
Suction Bottle
Capacity
60ml ( cc ) max.
Operation
Condition
Temperature : 41
˚
F - 95
˚
F / 5
˚
C - 35
˚
C
Relative humidity :30-85% RH
Storage
Condition
Temperature :-4
˚
F - 131
˚
F / -20
˚
C - 55
˚
C
Relative humidity :10-95% RH
S
AVE THESE INSTRUCTIONS
=Type BF
=class II
=Attention, consult accompanying documents
= Do not immerse in water
Tai Yu International Mfy. Ltd.
South Seas Center, Tower II, 8/F1., Unit 804 & 805A, 75 Mody Road, TST East, Kowloon, Hong Kong.
EU-representative:
P-W-Consult
Bitterstrasse 90, Koeln 50769, Germany
The MEDICAL DELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information provided in the ACCOMPANYING DOCUMENTS.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
PRODUCT FEATURES & SPECIFICATIONS
NOTE: The compressor uses an automatic reset thermal limiter that shuts off the
nebulizer if the motor is overheated due to misuse, abuse or abnormal operation.
The nebulizer will restart when the motor cools to an acceptable level.
NOTE: the Air Filter should be changed if significant dust is noticed around the middle of
the Air Filter.
NOTE: Alcohol is highly flammable; do not use alcohol in close vincinity of open fire; do
not smoke.