Emissions test
Compliance
Electromagnetic environment
guidance
WMPRF
emissions
CISPR 11
The model e-medic
TM
Silence
TP7 uses RF energy only for its
internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
The model e-medic
TM
Silence
TP7 is suitable for use in all es-
tablishments, including dome-
stic establishments and those
directly connected to the pu-
blic low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between
portable and mobile RF communications equipment and the
model
e-medic
TM
Silence TP7
The model e-medic
TM
Silence TP7 is intended for use in an electro-
magnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the model e-medic
TM
Silence
TP7 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the model e-medic
TM
Silence
TP7 as recommended below, according to the maximum output po-
wer of the communications equipment.
Rated maxi-
mum output
power of
transmitter
W
Separation distance according to
frequency of transmitter
m
150 kHz to
80 MHz
d = 1,2
80 MHz to
800 MHz
d = 1,2
800 MHz to
2.7 GHz
d = 2,3
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
8
Содержание e-medic SILENCE TP 7
Страница 1: ......
Страница 2: ...ii Version Change History Date Version Description Remark 2018 8 16 1 0 First release Cosa...
Страница 30: ...17 I O parts...
Страница 31: ...18 I O parts W ISO...
Страница 32: ...19 I O parts W USB 3 1 gen2...
Страница 35: ...22 Main...
Страница 47: ......
Страница 60: ...47 J16 H5 2mm J15 H9 2mm...