background image

9260

(A)

  9360

b3c

 pesage

pesage

w

ww.

b3

c-p

es

age

.fr

9260 (A) - 9360 V

er 2014 EN September BC

9260

(A)

9360

O

PERATING 

M

ANUAL 

(

GB

)

Compacts scales

Содержание Serenite 9260(A)

Страница 1: ...9260 A 9360 b3c pesage pesage w w w b 3 c p e s a g e f r 9260 A 9360 Ver 2014 EN September BC 9260 A 9360 OPERATING MANUAL GB Compacts scales...

Страница 2: ...ded for internal operations The scale complies with the type described in the EC type Approval Certificate as well as the appli cable requirement of EC Directives 89 336 EEC and 73 23 EEC SAFETY INSTR...

Страница 3: ...f there is permanent electrostatic interference we recommend to connect the platform with the ground The scale is a measuring instrument Air currents vibrations rapid changes in temperature and direct...

Страница 4: ...w When the transportation will be made please install the protection screw on the machine to avoid any damage sensitive components 2 LEVEL INDICATOR The scale is equipped with a level indicator on the...

Страница 5: ...an press this key the weight will become 0 Tare the weight of a container Indicate the net weight Input sample size To modify the backlight In the calibration mode to change the value 0 9 Accumulate w...

Страница 6: ...ht will be effective When the weight is zero for about 8 seconds the backlight will go off automatically bl 2 The backlight will always be on bl 3 The backlight will be turned off 6 TARE FUNCTION Tare...

Страница 7: ...ill show or pcs 9 ACCUMULATION FUNCTION In weighing or counting mode make it to zero place the goods 1 After the display is stable press the key to enter into accumulation mode 2 After accumulation to...

Страница 8: ...ble press the key Auto calibration is finished WITH AN KNOWN WEIGHT 1 Turn on scale 2 During the scale in self testing press the and keys about 3 seconds 3 When the display shows CC to change the figu...

Страница 9: ...ged If it the led is red the battery is nearly discharged If it the led is yellow the battery is increasing the charge level NOTE The battery should changed if the scale is not used for long time 12 T...

Страница 10: ...07 31 Load cell 15kg 0 2g 3 4 19020007 32 Load cell 15kg 0 5g 3 5 19020007 41 Load cell 30kg 0 5g 3 6 19020007 42 Load cell 30kg 1g 9 0210002 CPU Main board 10 0210001 LCD Display board 11 0210003 Key...

Страница 11: ...ENGLISH 11 3 1 3 2 3 3 3 4 3 5 3 6 1 2 9 10 11 12 3...

Страница 12: ...9260 A 9360 ES Balanzas electr nicas 9260 A 9360 IT Bilance electroniche 9260 A 9360 ND Electronische weegschaal Reeks 9260 A 9360 POR Balan as industriais 9260 A 9360 The indicated balances respect t...

Отзывы: