Instruction for Use
12
PRO-110 / PRO-115
En
8. REPLACEMENT OF THE AIR FILTER
The Air filter must be replaced in case of damage or severe contamination.
In normal usage conditions, the air filter must be replaced approximately after 500 work-
ing hours or after each year.
It is recommended periodically checking the air filter (10 – 12 treatments) and, if the filter
shows a grey or brown color or is wet, replace it.
1. Turn the air filter cover counterclockwise (only for PRO-115)
2. Extract the air filter bladder from the device by pulling it up
3. Extract the filter and replace it with a new one (for example with the help of tweezers)
4. Set the air filter bladder on its place
5. Turn the cover clockwise to fix it (only for the PRO-115)
WARNING!
The use of dirty filter or filter made of another material, for example,
cotton, may lead to device damage. Use original filters only.
•
Do not use the device without filter.
•
Do not try to clean the filter for reusing it. If the filter is wet replace it.
•
The air filter shall not be serviced or maintained while in use with a patient.
•
To avoid the air filter cover obstruction clean it regularly.
WARNING!
Air filters cannot be cleaned or washed.
9. CLEANING, MAINTENANCE AND STORAGE
Cleaning the device and the accessories
The cleaning of the device must be carried out by using a soft and dry cloth and non-
abrasive cleansers.
ATTENTION!
During cleaning operations, make sure the internal parts of the unit
are not in contact with liquids and that the power plug is disconnected.
Follow carefully the cleaning and disinfection instructions of the accessories as they are
very important for the device performances and the therapy success.
ATTENTION!
Before first using of the device and after each treatment it is
necessary to clean all the accessories according to the following instructions:
•
Disassemble the nebulizer by turning counterclockwise the top and remove the medi-
cine conduction cone.
•
Wash the components of the disassembled nebulizer, the mouthpiece and nosepiece
by using tap water; dip in boiling water for 5 minutes.
•
Reassemble the nebulizer components and connect it to the air-outlet, switch the
device on and let it work for 10-15 minutes or put on the clean paper towel for drying
on the air.
•
The masks and air tube must be washed with warm water. Does neither boil nor
autoclave the air tube and masks!
Sterilization:
•
Use cold disinfecting liquids following the instructions of the sterilization liquids’
manufacturer.
•
Wash your hands and take the components out from disinfection solution, wash the
components in warm running water, lay the components on paper towel and dry.
•
Do not wipe disinfected parts of the device with a towel; do not use hair dyer, micro-
wave oven or other household appliances for drying. The outer surface of wet parts
can be wiped with a clean dry cloth.
Caution:
In order to avoid spread of infections it is recommended to conduct disinfection
of device components
before first and
after every use. Use for treatment only the drugs
prescribed by your attending doctor and strictly follow his/her indications.
Opening of the device is prohibited. Repair of the device should be conducted only in
technical service centers recommended by B.Well company.
Note:
the nebulizer need to be replaced after 6-12 months depending on the usage.
The nebulizer must be replaced after a long period of inactivity, in case it shows
deformations or breakings, or when the nebulizer nozzle (diffuser) is obstructed by dry
medicine, dust, etc.
Use original nebulizers only.
Maintenance and storage
If you want your device to serve you for years, follow the following indications:
•
do not store the device at extremely high or low temperatures, high humidity or under
direct sunlight;
•
do not bend or wrap the air tube.
Operation and storage conditions – please see «12. Specification» section
Utilization of the device and any of its components should be performed in accordance
with the local regulations.
10. DISPOSAL
The unit must be disposed in accordance with current standards separately from
domestic wastes.
For disposal it is necessary to contact special organizations licensed to make utilization.
IM_PRO-110_PRO-115_EN_RO_BG_2916.indd 12
21.07.2016 11:39:03
Содержание PRO-110
Страница 22: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 22 21 07 2016 11 39 03...
Страница 44: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 44 21 07 2016 11 39 05...
Страница 47: ...Bg PRO 110 PRO 115 47 4 B Well PRO 115 M_PRO 110_PRO 115_EN_RO_BG_2916 indd 47 21 07 2016 11 39 05...
Страница 48: ...Bg PRO 110 PRO 115 48 9 16 8 ml 45 B Well M_PRO 110_PRO 115_EN_RO_BG_2916 indd 48 21 07 2016 11 39 05...
Страница 49: ...Bg PRO 110 PRO 115 49 11 5 1 2 3 4 5 6 2 6 3 1 5 M_PRO 110_PRO 115_EN_RO_BG_2916 indd 49 21 07 2016 11 39 06...
Страница 52: ...Bg PRO 110 PRO 115 52 6 9 1 2 3 4 7 M_PRO 110_PRO 115_EN_RO_BG_2916 indd 52 21 07 2016 11 39 07...
Страница 56: ...Bg PRO 110 PRO 115 56 12 10 11 30 30 8 ml B Well 12 M_PRO 110_PRO 115_EN_RO_BG_2916 indd 56 21 07 2016 11 39 07...
Страница 59: ...Bg PRO 110 PRO 115 59 5 A1 LOT M_PRO 110_PRO 115_EN_RO_BG_2916 indd 59 21 07 2016 11 39 07...
Страница 60: ...Bg PRO 110 PRO 115 60 15 BF II CE 0044 40 10 25 70 M_PRO 110_PRO 115_EN_RO_BG_2916 indd 60 21 07 2016 11 39 07...
Страница 67: ...M_PRO 110_PRO 115_EN_RO_BG_2916 indd 67 21 07 2016 11 39 08...
Страница 68: ...IM_PRO 110_PRO 115_EN_RO_BG_2916 www bwell swiss ch M_PRO 110_PRO 115_EN_RO_BG_2916 indd 68 21 07 2016 11 39 08...