Technical Specifications
Product Name
B.O.L.T
Model
VA 01
Base Unit
Base Unit Dimensions
130 (L) X 40.3 (H) X 75 (D) mm
Base Unit Weight
220 grams (Including Battery)
Power Source
DC 5V 2A
Battery
3.7V Li-Polymer 850mAH
Bluetooth
Bluetooth 4.0 EDR or Lower, Bluetooth Low Energy (BLE)
Non-Invasive Blood Pressure
Blood pressure measurement method
Oscillometric method with automatic inflation &
measurement
Cuff Pressure range
0 to 280 mmHg
Measurement Pressure
SYS: 40 to 260mmHg
DIA: 25 to 220mmHg
Pressure Accuracy
±3mmHg
BP Pulse range
30 to 200 BPM
Pulse Accuracy
± 5%
Cuff Limb Circumference
22 cm to 42 cm
Cuff Size
610 mm * 145 mm
Cuff Weight
205 Grams
IR Thermometer (IRT) Probe
Body Temperature measurement
method
Tympanic, Infrared based
Body temperature display range
34 °C to 43 °C (93.2 °F to 109.4 °F)
Measuring site
Ear
Body temperature Accuracy
-
(from 36°C to 39.0°C) / (from
96.8°F to 102.2°F)
-
outside this range
±0.2 °C / ± 0.4°F
±0.3 °C / ± 0.5°F
IRT Display resolution temperature range
0.1 °F or °C
IRT Unit Dimensions
84.8 (L) X 18 (H) X 26.6 (D) mm (Highest width along the
profile)
IRT Unit Weight
42 grams
Pulse Oximeter (SpO2) Probe
SpO2 measurement method
Optical method
SPO2 range
70% to 100%, below 70%: Unspecified
SpO2 Measurement Accuracy
70% to 100% - ±2%, below 70%: Unspecified
SpO2 Pulse Rate Range
30 to 180 BPM
SpO2 Pulse Accuracy
± 2%
SpO2 Unit Dimensions
58.2 (L) X 35 (H) X 35.3 (D) mm
SpO2 Unit Weight
60 grams