![B meters Hydrocal-M3 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/b-meters/hydrocal-m3/hydrocal-m3_user-manual_3068775026.webp)
Hydrocal M3 User Manual v1.8
ENG - 26
Model
Hydrocal M3
Power supply
Battery supply
Battery type
Li-SoCl2 Lithium-thionyl chloride, 3,6V 6Ah
Battery Life
10 years +1 (depending on the version)
Operating temperature range
5 ÷ 55°C
Storage temperature range
-10 ÷ 55°C
Dimensions
112 x 78 x 76.5mm
Weights
635g (DN15); 708g (DN20)
IP protection grade
IP54
Approval
2004/22/CE MID (Module B)
Certificate nr. IT-002-15-MI004-NB2213
EN1434
Environmental class
A (E1, M1)
Temperature measurement range, MID
approved (heating applications)
Θ: 5÷90°C
Temperature difference range, MID
approved (heating applications)
ΔΘ: 3÷70K
Temperature measurement range
(cooling)
Θ: 0.2÷24°C
Temperature difference range
(cooling)
ΔΘ: 0.2÷20K
Accuracy class
2
Range
Depending on diameter and nominal flow rate
DIAMETER
Qp (
m
3
/h)
Range (H/V)
DN15
0.6
25/25
DN15
1.5
50/50
DN20
2.5
50/50
Operative counting conditions
Heating: ΔΘ≥1K and fluid temperature ≥5°C (accounting enabling
conditions)
Cooling: ΔΘ≤0.2K and fluid temperature ≤24°C
Maximum measurable power
650 kW
Display
LCD, 8 icons
Measurement unit
MWh (standard), GJ (optional)
Temperature sensors
Pt1000
Cable length
1.5 m
Pulse inputs
2 dedicated for the sanitary water volume meters
Pulse input class
Class IA (default): Open Collector or electromagnetic contact (reed).
Pulse input maximum frequency
5 Hz
Maximum length pulse emitter cable
3 m
Installation site of flowmeter
Return flow pipe (Default), Inlet flow pipe (optional).
Always refer to the indication on the cover of the meter.
Supported vector fluid
Water