Mod.BM43
COM
Vers.1.0
16/11/2005
Page 54 of
95
B.M. Tecnologie Industriali
Via Praimbole 13
35010 LIMENA (PD)
VAT No. IT02459940280
Tel. +39 (0) 49 884.16.51
Fax +39 (0) 49 884.16.54
E-Mail [email protected]
Web www.bmtecnologie.it
Digit
and the device will display
M 6 1
V e r 8 . 3 8 x 3 2
E S N = 0 6 0 8 0 2 6 2 H
This window displays the software version and the serial number of the device (ESN). Due
to continuous software updating, the software versions could be different.
Digit
and the device will display
M 6 2
R S - 2 3 2 C S e t u p
B a u d r a t e : = 1 9 2 0 0 ,
P a r i t y :
N o n e
This window is used to set up the communication parameters of the serial port. Press ENT
and then use the arrow keys to select among:
Baud Rate:
115200
57600
38400
14400
28800
19200
9600
4800
3600
2400
1200
600
300
150
110
75
Press ENT to confirm the selected option and then select the parity type among:
None
Even
Odd
Press ENT to confirm the selection and exit the setting.
The other parameters are factory set up: 8 bit data, 1 bit stop and flow control Xon/Xoff.