7.2
Program Settings
Press the Program softkey in the Setting and Measurement page to enter Program
mode. There are three modes for the user to choose from to simulate Power Line
Disturbances. They are:
Step Output settings step-up or step-down based on user criteria.
List
Change the output sequentially by individual settings included in a list.
Pulse Trigger pulse signals periodically.
7.2.1
Step Mode
Step Mode has 10 parameters that can be set by the user. The following table describes
each parameter with its range and description.
Parameters
Range
Description
Volts (rms)
0 to 300.0 Vrms
Initial AC output voltage
dVac
0 to ±300.0 Vrms
In/decrement of VAC per
step
Volts (DC)
-424.0 to 424.0 V
Initial DC output voltage
dVdc
-424.0 to 424.0 V
Increment/decrement of
VDC per step
Freq.
43 to 1200.0 Hz
Initial output frequency
dF
±43 to 1200.0 Hz
Increment/decrement of
FREQ. per step
Set Time
0 to 100000 ms
Set the time interval of
one step
Count
1 to 99
Set how many steps to
run
Sync Source
Immed, Phase
Select the mode of
output transient phase
Phase (deg)
0.0
゜
to 359.7
゜
Set the angle of output
transient
Table 2 - Step Mode Parameters
Содержание 9833B
Страница 1: ...1 9830B Series USER MANUAL Model 9832B 9833B High Power Programmable AC Power Sources ...
Страница 10: ...ix 2 Notations TEXT Denotes a softkey TEXT Denotes a front panel key ...
Страница 28: ...15 6 Menu Tree ...
Страница 86: ...73 THD 11 Harmonic 3 weight 21 3 THD 12 Harmonic 3 weight 24 5 ...
Страница 87: ...74 THD 13 Harmonic 2 5 7 8 weight 2 3 9 8 15 8 2 5 THD 14 Harmonic 2 5 7 8 weight 1 15 4 9 7 9 1 25 ...
Страница 88: ...75 THD 15 Harmonic 5 7 weight 2 45 3 95 THD 16 Harmonic 3 5 7 9 weight 11 4 05 2 1 3 ...
Страница 89: ...76 THD 17 Harmonic 3 5 9 weight 7 17 3 42 0 8 THD 18 Harmonic 3 5 9 weight 8 11 3 48 1 ...
Страница 90: ...77 THD 19 Harmonic 3 5 9 weight 9 38 3 44 1 15 THD 20 Harmonic 3 5 7 9 11 weight 2 1 8 1 6 1 23 0 9 ...
Страница 95: ...82 THD 29 Harmonic 3 5 7 9 11 13 15 weight 33 3 20 13 8 10 8 8 5 7 2 5 5 ...
Страница 103: ...90 1 The instrument will show Cal End then press ESC to complete the external voltage calibration ...
Страница 108: ...95 22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2020 B K Precision Corp Printed in Taiwan v101320 ...